manuals for the connecting instruments.
3.
Without repeaters, the RS-485 network cannot exceed a maximum of 32 nodes. The maximum
communication transmission distance of RS-485 is up to 1200 meters. For long distance
transmission, it is recommended to apply cables which are dedicatedly design for RS-485.
4.
When communication, all the equipments of the network should be maintained in the same transfer
mode, baud rate, parity consistent. And each of the device address cannot be the same, so as not to
conflict resulted in the normal network communications.
5.
The Modbus command of the transmitter EC-4110-I/EC-4110-ICON can only access 50 registers.
If it exceeds the upper limit, then it returns abnormal message.
6.
The waiting time which a slave instrument responses to a master machine is different according
to each model. Generally, it shall be longer than 0.5 second. (Some models may require a longer
waiting-responding time, please note whether it is specified in the operation manual.)