Cal Type: No Cal
Cal Temperature: None
Cell Constant: 2.700
Zero: 0.0
ȝ
S
Auto back: Auto, 3 minutes
Code set-up: Off
Note:
The factory default of calibration presetting is “No Cal”, and the cell constant setting is “2.700”. It means
that the user has not calibrated the sensor with the transmitter yet. When selecting standard solution or
directly adjusting cell constant to finish calibration, the display shows cell constant of the cell. It should
re-start a zero-point calibration in air after installation or maintenance, and the surface of sensor needs
cleaning without attachment with cleaning solutions during calibration.