7.8 Relay 1
Enter setup of Relay 1. Select the item to turn on or turn of the relay 1 function. If you select to
turn on the relay 1, then select for using relay 1 as “High set-point” alarm or “Low set-point”
alarm. Set the value of set-point (SP) and Hysteresis (Hys.). The relationship between parameters
can refer to an explanatory diagram of the box (as a high point alarm).
Press to confirm it.
Press to confirm it.
Press to confirm it.
Press to confirm it.
Press to confirm it.
Press to confirm it.
Press to confirm it.
Press to confirm it.
High Point Alarm Control
off off
on
on
SP
Hys.
%/mS/
ȫ
S/ppm/ppt
t (sec)
ņůŵŦųȾ
Relay 2
ȿŔŦŵŶűġ
Select
ġ
AUTO.
ġ
Press to confirm it.
Press to confirm it.
Press to confirm it.
Press
ġ
or
to select Hi point
or Lo point alarm.
Press or
to adjust Hysteresis
(Hys.) value
Press or to adjust
set-point (SP) value.
If the measuring range of
conductivity is set as AUTO,
then it needs to be set unit
ȝ6FPRUP6FPYDO
ue, and
decimal point by orders.
Press
ġ
or to
select to activate a test on
REL1. If “ON”, the relay
1 is in action, and the
“ACT” indication light
will light up.