SPECIFICATIONS
13
F-20L Compressor
There are two electrical voltage variations:
Compressor P/N
Mains Voltage
280090D18N
208-230 VAC +/- 10%, V~, 60 Hz and
200 VAC +/- 10%, V~ 50 Hz
280090D27N
220 - 240 VAC +/- 10%, V~ 50 Hz
Electrical Characteristics
Service required: 2 poles, 3 wires (single phase plus protective ground.)
Mains Power Receptacle F-20L: Phoenix Heavycon Advance – Type HC-B10-TMS with screw
locks.
Power Consumption
Variation Frequency
[Hz]
Voltage Operation
[kW]
Compressor
RLA [A]
Compressor
LRA [A]
MCA
[A]
MOP
[A]
18N
60
220
2.6
12.0
56
15.0
30.0
50
200
2.4
12.0
15.0
30.0
27N
50
230
2.25
10.0
43-47
13.0
25.0
RLC = rated load current
MCA = minimum circuit ampacity
LRA = locked rotor amps
MOP = maximum overcurrent protection
Mains Power Receptacle: Phoenix Heavycon Advance – Type HC-B10-TMS with screw locks
Mains Power Supply Cable: 12 AWG, 3 conductor, 300V cordage with Phoenix connector and
NEMA L6-20P (2 pole, 3 wire, 20A, 250 V~) locking plug, P/N 280112C10
Circuit Protection:
Input: 20A, 250V circuit breaker
Control Circuit: (2) FU1 & FU2, 1.0A Slo Blo, 5 x 20 m fuses
Cold Head Valve Motor Circuit: (2) FU3 & FU4, 1.0A Slo Blo, 5 x 20 mm fuses
All fuses are located on front control panel
Содержание F-20L
Страница 2: ......
Страница 10: ...6 This page is intentionally blank...
Страница 12: ...8 This page is intentionally blank...
Страница 23: ...Specifications 19 Figure 2 F 20L Compressor Front View...
Страница 24: ...Specifications 20 Figure 3 F 20L Compressor Dimensions Dimensions are in inches and mm...
Страница 34: ...30 This page is intentionally blank...
Страница 36: ...32 This page is intentionally blank...
Страница 48: ...44 This page is intentionally blank...
Страница 53: ...Troubleshooting 49 Figure 19 F 20L Compressor Wiring Diagram...
Страница 54: ...Troubleshooting 50 Figure 20 F 20L Compressor Wiring Schematic...