6206-001-700 Rev-
2006/05
www.stryker.com
9
To Drain Water From Handpiece
To Clean Handpiece
Remove the battery pack and cutting accessory from
the handpiece.
Using a brush with stiff, non-metallic bristles and
hospital enzymatic cleaner, scrub the debris from
the handpiece. Pay special attention to crevices and
other hard to reach areas such as seams, joints, and
details around the blade retainer, trigger, and con-
nector areas.
Rinse all the external surfaces of the handpiece
under running water. Hold the handpiece upright to
prevent water from running into the battery pack
contact area.
If water leaks into the handpiece, tip the handpiece
back as shown below to allow drainage from a small
opening in the battery pack contact area.
1.
2.
3.
4.
Visually inspect the handpieces for any remaining
debris; if any debris is present, repeat the cleaning
and rinsing procedure using fresh hospital enzymatic
cleaner.
Dry the handpiece with a lint-free towel.
After cleaning, sterilize as directed. See
Sterilization
Recommendations
.
5.
6.
7.
Cleaning Recommendations
WARNINGS:
• Clean and sterilize handpieces and batteries before
first and every use.
• Prior to cleaning and sterilization, remove all acces-
sories from the handpieces.
• DO NOT use solvents, lubricants, or other chemicals
unless otherwise specified.
CAUTIONS:
• DO NOT immerse a handpiece or battery pack in
liquid. Moisture may enter the equipment, cause cor-
rosion, and damage the electrical and/or mechanical
components.
• DO NOT allow liquid to run directly into any electrical
connection. Moisture may cause corrosion to electri-
cal components.
To Clean Battery Packs and Accessories
See the care instructions supplied with the battery
packs, battery pack modules and battery charger.