Troubleshooting Guide*
PROBLEM
CAUSE
ACTION
Handpiece does not run or re-
ciprocates at a reduced speed
making cutting difficult.
Battery pack is discharged.
Recharge the battery pack in Stryker charger.
Battery pack is expended.
Replace the battery pack.
Run/safe control is in the safe
position.
Slide the run/safe control to the run position.
Drivetrain is malfunctioning.
Return the handpiece for repair.
Motor runs but blade does not
move.
Drivetrain is malfunctioning.
Return the handpiece for repair.
Battery pack becomes unusually
hot during use.
Circuitry is malfunctioning.
Check the battery pack on the charger. Replace the bat-
tery pack if required. See the instructions supplied with
the battery charger.
Blade will not fit into the blade
collar.
Debris is inside the end of the
blade collar.
Clean the handpiece with a small brush.
Blade collar is damaged.
Return the handpiece for repair.
Blade collar does not spring
back to retain the blade.
Debris inside of blade collar.
Clean the handpiece with a small brush. Actuate the
blade collar several times to obtain smooth operation.
Handpiece has become noisy
and vibrates.
Drivetrain is malfunctioning.
Return the handpiece for repair.
Sporadic electrical interference
is experienced.
electrical noise is present.
Turn off all electrical equipment not in use in the operat-
ing room.
Relocate electrical equipment; increase spatial distance.
Plug operating room equipment into different operating
room outlets.