26
Troubleshooting
Note
If a fault condition should occur, the display on the front panel
will display an error code and an error beep will sound. The
interpretation of the error code, as well as possible causes and
solutions are listed in Table 1.
Note
To recover system from Errors P1-P4, E1-E9, press any button or
re-connect the Probe to the Generator. If errors persist, refer to
Solution section in Table 1. If solutions do not prevent error
occurrence, refer to the Service and Claims section of this manual
to obtain repair service.
Table 1:
RECOVERABLE ERRORS & CORRESPONDING CODES
Error Code
Reason for Error
Solutions
P1
Probe: Probe not ready
•
Check Probe connection to
Generator. Reconnect Probe.
P2
Probe: Probe expired
•
Discard Probe and replace with
new Probe
P3
Probe: Probe identification invalid
•
Replace Probe
P4
Probe: Probe communication error
•
Check Probe connection to
Generator. Reconnect Probe.
•
Replace Probe
E1
Unable to identify error
•
Check Probe for damage
•
Replace Probe
E2
RF Power too high
•
Check Probe for damage
•
Replace Probe
E3
RF Voltage too high
•
Check Probe for damage
•
Replace Probe
E4
RF Current too high
•
Check Probe for damage
•
Replace Probe
E5
RF delivery exceed countinous use limit •
Clear error and continue
E6
Generator temperature too high
•
Ensure proper airflow on sides
and rear of Generator
E7
Low impedance detected
•
Check Probe for damage
•
Replace Probe
E8
Switch pressed upon Probe/Footswitch
correction
•
Check that switches on Probe,
Footswitch, and Generator are
not pressed
Содержание SERFAS
Страница 2: ......
Страница 10: ...8 ...
Страница 11: ...SERFASEnergy System Operation and Maintenance Manual Endoscopy Energy System ...
Страница 43: ...SERFASEnergy System Bedienungs und Wartungshandbuch Endoscopy Energy System ...
Страница 75: ...73 ...
Страница 77: ...SERFASSistema Energy Manual de utilización y mantenimiento Endoscopy Sistema Energy ...
Страница 111: ...SERFASSystème Energy Manuel d utilisation et de entretien Endoscopy Système Energy ...
Страница 143: ...141 ...
Страница 145: ...SERFASSISTEMA Energy Manuale di funzionamento e manutenzione Endoscopy SISTEMA Energy ...
Страница 179: ...SERFASSistema Energy Manual de Operação e Manutenção Endoscopy Sistema Energy ...
Страница 211: ...209 ...
Страница 213: ...SERFASEnergy systeem Gebruiks en onderhoudshandleiding Endoscopy Energy systeem ...
Страница 247: ...SERFASEnergy system Bruger og vedligeholdelsesvejledning Endoscopy Energy system ...
Страница 279: ...SERFASEnergy järjestelmä Käyttö ja huolto opas Endoscopy Energy järjestelmä ...
Страница 311: ...SERFASEnergy System Manual för drift och underhåll Endoscopy Energy System ...
Страница 343: ...SERFASȈȪıIJȘµĮ Energy ǼȖȤİȚȡȓįȚȠ ȜİȚIJȠȣȡȖȓĮȢ țĮȚ ıȣȞIJȒȡȘıȘȢ Endoscopy ȈȪıIJȘµĮ Energy ...
Страница 377: ...SERFASEnergy System Podręcznik użytkowania i obsługi Endoscopy Energy System ...
Страница 409: ...407 ...
Страница 411: ...SERFAS Energy ΏΑΞθ ե ྶ Endoscopy Energy ΏΑΞθ ...
Страница 443: ...SERFAS øœµÕŠ øœµÕŠ Ϣ㓈ᡸ ݠ Endoscopy 能量系统 ...
Страница 467: ...SERFAS ø Šþ ˆ Ð Ý ø Šþ ˆ Ð Ý 㧧 G G㥶㰖ὖⰂG ⳛ GGG Endoscopy 櫖嘎滆 柢枪癢 ...
Страница 493: ...491 ...