5. Using a 5/32” hex wrench, remove the two button head cap screws (E) on the bottom of the go pedal. Save the screws.
When reinstalling the screws, use a 3/8” drive torque wrench to torque the screws to 70 ± 10 in-lb.
M
N
P
R
T
U
V
E
F
G
H
K
L
J
J
F
A
B
C
D
Torque item M
to 70 ± 10 in-lb
F
Fiig
gu
urre
e 1
12
2 –– P
Pe
ed
da
all rre
ep
plla
acce
em
me
en
ntt
6. Using needle nose pliers, remove the rue ring cotter (F) on both ends of the toggle link rod (G). Save the rue ring cotters.
7. Remove the toggle link rod (G).
8. Using a 5/32” hex wrench and a 3/16” hex wrench, loosen the button head cap screw (H) that secures the brake cam
spring pad (K) to the chair frame.
N
No
otte
e -- Do not remove the brake finger spacer nut (J).
9. Use your hand to push on the brake cam spring pad (K) to relieve pressure from the brake finger spacer nut (J). Remove
and save the brake finger spacer nut and screw.
10. Carefully remove the brake cam spring pad (K) and brake cam spring (L). Save the pad and spring.
11. Repeat steps 8-10 for the other side.