![Stryker Prime TC Скачать руководство пользователя страница 18](http://html.mh-extra.com/html/stryker/prime-tc/prime-tc_maintenance-manual_1386000018.webp)
A
B
C
Torque item A
to 33 ± 7 ft-lb
F
Fiig
gu
urre
e 7
7 –– F
Flliip
p--u
up
p a
arrm
mrre
esstt rre
ep
plla
acce
em
me
en
ntt
5. Pulling upward on the armrest release pin, install the supplied flip-up armrest onto the chair frame.
6. Using a 5/16” hex wrench, install the supplied hard stop pin (removed in step 2). Using a 3/8” drive torque wrench,
torque the hard stop pin to 33 ± 7 ft-lb.
7. Verify proper operation before you return the product to service.
F
Frro
on
ntt cca
asstte
err w
wh
he
ee
ell rre
ep
plla
acce
em
me
en
ntt
T
To
oo
ollss rre
eq
qu
uiirre
ed
d::
•
Pick
•
(2) 5/32” hex wrench
•
3/8” drive torque wrench (in-lb)
P
Prro
occe
ed
du
urre
e::
1. Position the chair on its back.
2. Using a pick, remove and save the caster wheel axle barrel nut covers (C) (Figure 8).
3. Using two 5/32” hex wrenches, remove the axle barrel nut (B) and axle screw (D). Save the axle barrel nut and axle
screw. When reinstalling the axle screw, use a 3/8” drive torque wrench to torque the axle screw (D) to 70 ± 10 in-lb.
4. Grasp the caster wheel (A) and pull outward to remove the wheel from the fork (E). Discard the wheel.
EN
14
1460-009-002 Rev B.0
Содержание Prime TC
Страница 2: ......
Страница 4: ......
Страница 38: ...Torque item B to 70 10 in lb Torque item B to 70 10 in lb EN 34 1460 009 002 Rev B 0 ...
Страница 63: ......
Страница 64: ...Stryker Medical 3800 E Centre Avenue Portage MI 49002 USA 1460 009 002 Rev B 0 WCR AA 5 2019 09 ...