![Stryker Prime TC Скачать руководство пользователя страница 16](http://html.mh-extra.com/html/stryker/prime-tc/prime-tc_maintenance-manual_1386000016.webp)
4. Push back on the chart holder from the bottom to remove the chart holder from the lower chair frame, and then push up
to remove the chart holder from the upper chair frame.
A
Torque item A
to 25 ± 5 in-lb
F
Fiig
gu
urre
e 5
5 –– C
Ch
ha
arrtt h
ho
olld
de
err rre
ep
plla
acce
em
me
en
ntt
5. Reverse steps 2-4 to install the supplied chart holder.
6. Verify proper operation before you return the product to service.
S
Se
ea
att rre
ep
plla
acce
em
me
en
ntt
T
To
oo
ollss rre
eq
qu
uiirre
ed
d::
•
3/16” hex wrench
•
3/8” drive torque wrench (in-lb)
P
Prro
occe
ed
du
urre
e::
1. Using a 3/16” hex wrench, remove the four button head cap screws (A) that secure the seat (B) to the chair frame
(Figure 6). Discard the seat and screws.
2. Using a 3/16” hex wrench, install the four supplied button head cap screws (A) to secure the supplied seat (B) to the
chair frame. Using a 3/8” drive torque wrench, torque the screws to 70 ± 10 in-lb.