Reprocessed LigaSure Impact™ Curved, Large Jaw, Open Sealer/Dividers Description 

The Reprocessed LigaSure Impact

TM

 Curved, Large Jaw, Open Sealer / Divider (LF4318) is a hand-held bipolar electrosurgical 

instrument designed for use with Covidien electrosurgical generators that include vessel sealing capability. Please refer to the 
cover page for details on compatible generator models and software versions. If the software version on your generator is 
lower than required, contact Covidien about software upgrades. 

 
The instrument has a shaft diameter of 13.5 mm (square), shaft length of 18 cm, and a jaw length of 36 mm.  The following 
controls are located on the instrument handle: 

 

A lever for opening and closing the instrument jaws. The mechanism incorporates a latch to hold the jaws in the 

closed position during vessel sealing and cutting. 

 

An activation button for generator power to initiate vessel sealing. 

 

A trigger for actuating the cutter.  

 

A knob to rotate the instrument jaws. 

 
All controls can be operated with either the right or left hand.  Vessel sealing can be initiated using the activation button or 
utilizing a footswitch connected to the generator. 

 
Maximum rated voltage: 288 V

peak

 

 

Indications for Use 

The Reprocessed LigaSure Impact

TM

 Curved, Large Jaw, Open Sealer / Divider (LF4318) is a dedicated bipolar electrosurgical 

instrument intended for use in open surgical procedures where ligation and division of vessels is desired.  The LF4318 is 
intended to be used with the Covidien electrosurgical generators that include vessel sealing capability to cut and seal vessels, 
and to cut, grasp, and dissect tissue during surgery.  The indications for use include open procedures (general, urologic, 

vascular, thoracic, and gynecological) where ligation and division of vessels is performed.  These procedures include vaginal 
hysterectomies, Nissen fundoplication, colectomy, adhesiolysis, oophorectomy, etc.  The LF4318 can be used on vessels 
(arteries, veins, pulmonary vasculature, and lymph) up to and including 7 mm and tissue bundles. 

 

The LigaSure

TM

 system has not been shown to be effective for tubal sterilization or tubal coagulation for sterilization 

procedures.  Do not use the LigaSure

TM

 system for these procedures. 

 

  

 

 

Contraindications for Use 

Reprocessed 

LigaSure Impact™ Curved, Large Jaw, Open Sealer/Dividers Description

 are contraindicated for:  

 

None 

 

Warnings 

 

The LF4318 instrument is intended for use ONLY with the equipment listed on the cover of this document.  Use of this 

instrument with other generators may not result in the desired tissue effect, may result in injury to the patient or surgical 
team, or may cause damage to the instrument. 

 

Do not use the LigaSure System unless properly trained to use it in the specific procedure being undertaken.  Use of this 

equipment without such training may result in serious unintended patient injury. 

 

Use caution during surgical cases in which patients exhibit certain types of vascular pathology (atherosclerosis, 

aneurysmal vessels, etc.).  For best results, apply the seal to unaffected vasculature. 

 

Do not use this instrument on vessels in excess of 7 mm in diameter. 

 

These instruments are only intended for use by individuals with adequate training and familiarity with the specific 

procedure being undertaken.  Use of this equipment without such training may result in serious unintended patient injury.  
For further information about techniques, complications and hazards, consult the medical literature. 

 

Package is provided sterile by method of ethylene oxide gas and is for single patient use only.  Do not use if there is any 

evidence of damage to the package. 
 

 
 
 

Содержание LF4318

Страница 1: ...ILE NOT MADE WITH NATURAL RUBBER LATEX Explanation of Symbols Sterilized by Ethylene Oxide Gas Date of Reprocessing Use by Date Do Not Reuse See Instructions For Use Stryker Sustainability Solutions Inc 2017 1810 W Drake Dr Tempe AZ 85283 sustainability stryker com 888 888 3433 Compatible with LF4318 FORCETRIAD SW v3 5 v4 0 VLFT10GEN SW v1 1 v2 0 1 13 NOTE Devices with a Valleylab Plug are NOT com...

Страница 2: ...lude open procedures general urologic vascular thoracic and gynecological where ligation and division of vessels is performed These procedures include vaginal hysterectomies Nissen fundoplication colectomy adhesiolysis oophorectomy etc The LF4318 can be used on vessels arteries veins pulmonary vasculature and lymph up to and including 7 mm and tissue bundles The LigaSureTM system has not been show...

Страница 3: ...damage before use Failure to observe this caution may result in injury or electrical shock to the patient or surgical team or cause damage to the instrument If damaged do not use Position instrument cords to avoid contact with the patient or other leads Do not wrap cords around metal objects This may induce currents that can lead to shocks fires or injury to the patient or surgical team Ensure all...

Страница 4: ...s this may result in an incomplete seal If the seal cycle complete tone has not sounded an optimal seal may not have been reached Reactivate the RF energy until a seal complete tone is heard Caution Pediatric applications and or procedures performed on small anatomic structures may require reduced power settings The higher the current flow and the longer the current is applied the greater the poss...

Страница 5: ...nt jaws with a scratch pad Troubleshooting Alert Situations Note If both the activation button and footswitch pedal are activated during the same seal cycle the vessel sealing system delivers energy from the activation source it detects first A tone with multiple pulses indicates that the seal cycle was not completed Do not cut tissue until you have verified that there is an adequate seal When an ...

Страница 6: ...the seal cycle was complete After Surgery Discard the LF4318 instrument after use according to hospital policy for biohazards and sharps Pre Clinical Study Notice There is no animal data qualified to predict the effectiveness of this device in sealing vessels containing atherosclerotic plaque Product performance of the device was established in a chronic in vivo porcine model The results showed th...

Страница 7: ...assigned without the express written consent of Stryker This warranty does not apply to 1 products that have been misused neglected modified altered adjusted tampered with improperly installed or refurbished 2 products that have been repaired by any person other than Stryker personnel without the prior written consent of Stryker 3 products that have been subjected to unusual stress or have not bee...

Страница 8: ...third parties that do not sponsor this device Sterilization This product and its packaging have been sterilized with ethylene oxide gas EtO Even though the product then is processed in compliance with all applicable laws and regulations relating to EtO exposure Proposition 65 a State of California voter initiative requires the following notice Warning This product and its packaging have been steri...

Страница 9: ...IQUÉ À PARTIR DE LATEX DE CAOUTCHOUC NATUREL Explication des symboles Stérilisé par oxyde d éthylène Date du retraitement Date de péremption Ne pas réutiliser Voir le mode d emploi Compatible avec LF4318 FORCETRIAD SW v3 5 v4 0 VLFT10GEN SW v1 1 v2 0 1 13 REMARQUE Les appareils avec une prise Valleylab ne sont PAS compatibles avec le modèle VLFT10GEN SW v1 1 5 4 ou ultérieur Stryker Sustainability...

Страница 10: ...terventions chirurgicales ouvertes générale urologique vasculaire thoracique et gynécologique dans lesquelles la ligature et la division des vaisseaux sont effectuées Ces interventions comprennent entre autres les hystérectomies vaginales la fundoplication la colectomie l adhésiolyse et l oophorectomie Le LF4318 peut être utilisé sur les vaisseaux artères veines vaisseaux pulmonaires et lymphatiqu...

Страница 11: ...ation inspectez l instrument et les cordons pour déceler toute trace de rupture de fissure d entaille ou de tout autre Le défaut d observer cette mise en garde pourrait entraîner une blessure ou une décharge électrique au patient ou à l équipe chirurgicale ou provoquer des dommages à l instrument Si l instrument est endommagé ne l utilisez pas Positionnez les cordons de l instrument de façon à évi...

Страница 12: ...vez pas l énergie RF jusqu à ce que la tige ait été correctement verrouillée L activation du dispositif avant cette étape pourrait entraîner une étanchéisation inadéquate et pourrait augmenter la diffusion thermique à des tissus à l extérieur du champ opératoire N activez pas l instrument lorsqu il est en contact avec d autres instruments ou à proximité incluant les canules Des brûlures localisées...

Страница 13: ...ostase visible Si le scellement n est pas adéquat répétez le cycle de scellement en ajustant les paramètres au besoin 2 Pour introduire le mécanisme de coupe tirez complètement sur le déclencheur de coupe vers l arrière en direction du corps de l instrument La lame de l instrument coupe le scellement d environ 2 mm à partir de l extrémité distale de la plaque de scellement 3 Ouvrez les mâchoires e...

Страница 14: ...ule autour de l embout de l instrument Alerte Réactivation cycle de scellement incomplet Une condition d alerte de réactivation cycle de scellement incomplet produit une séquence de tonalités pulsées et une alerte s affiche sur le générateur Si l alerte de réactivation cycle de scellement incomplet s affiche l utilisateur doit 1 Relâchez la pédale ou le bouton d activation 2 Réactiver le cycle de ...

Страница 15: ... applicables à tous les produits DANS TOUTE LA MESURE PERMISE PAR LA LOI LA GARANTIE EXPRESSE MENTIONNÉE AUX PRÉSENTES EST LA SEULE GARANTIE APPLICABLE AUX PRODUITS ET REMPLACE EXPRESSÉMENT TOUTE AUTRE GARANTIE DE STRYKER EXPRESSE OU IMPLICITE INCLUANT MAIS SANS S Y LIMITER TOUTE GARANTIE IMPLICITE OU DE QUALITÉ MARCHANDE OU D ADÉQUATION À DES FINS PARTICULIÈRES EN AUCUN CAS LA RESPONSABILITÉ DE S...

Страница 16: ... ou si la période de la garantie initiale est expirée au moment où le produit est réparé ou remplacé pour une période de trente 30 jours après la livraison du produit réparé ou remplacé Lorsqu un produit ou un composant est remplacé l article fourni en remplacement deviendra la propriété du client et l article remplacé la propriété de Stryker Si un remboursement est fourni par Stryker le produit p...

Отзывы: