Electromagnetic Compatibility
Like other electrical medical equipment, the L10 AIM Light Source requires
special precautions to ensure electromagnetic compatibility with other
electrical medical devices. To ensure electromagnetic compatibility (EMC), the
L10 AIM Light Source must be installed and operated according to the EMC
information provided in this manual.
Note:
The L10 AIM Light Source has been designed and tested to comply
with IEC 60601-1-2 requirements for EMC with other devices.
Note:
This equipment is for use in a professional healthcare environment. It is
not for use in the RF shielded room of a medical electrical system for magnetic
resonance imaging, where the intensity of EM disturbances is high.
Note:
This equipment is not likely susceptible to interference from HF
surgical instruments in the Special Environment of being in close proximity
to an active HF surgical instrument. In the case that HF surgical interference is
observed, adjust the separation distance of the equipment.
Do not use cables or accessories other than those provided
with the L10 AIM Light Source, as this may result in increased
electromagnetic emissions or decreased immunity to such
emissions.
If the L10 AIM Light Source is used adjacent to or stacked with
other equipment, observe and verify normal operation of the
L10 AIM Light Source in the configuration in which it will be
used prior to using it in a surgical procedure. Consult the tables
below for guidance in placing the L10 AIM Light Source.
Equipment which employs RF communications may affect the
normal function of the L10 AIM Light Source.