26
Required Supplies
AIM IC-GREEN® (NDC # 17478-801-02) includes the required ICG supplies.
Other items required to use AIM IC-GREEN shall be provided by the user.
Note:
In the United States, the L10 LED Light Source with AIM Technology
and SafeLight Fiber Optic Cable are indicated for use only with AIM IC-GREEN.
For users outside the U.S., refer to Stryker user manual P27009.
Item
Description
AIM IC-GREEN
Includes six 25 mg vials of ICG powder and
six 10 mL ampules of Sterile Water for
Injection, USP
Sterile normal saline
Approximately 10 –12 mL following each ICG
injection
Syringe
(for ICG preparation)
Minimum volume of syringe: 10 mL
Syringes
(for each ICG solution
injection)
Minimum volume of syringes: 3 or 5 mL
Syringes
(for each saline flush
after ICG injection)
Minimum volume of syringes: 12 mL
ICG Preparation
Prepare the ICG solution following the steps below. The instructions yield a
2.5 mg/mL solution.
1. Open the Sterile Water for Injection, USP ampule.
2. Draw 10 mL of the solvent into a syringe.
3. Remove the flip cap off the vial containing 25 mg of the ICG powder.
4. Inject the 10 mL of Sterile Water for Injection, USP into the ICG vial
through the stopper to dissolve the powder.
5. Gently shake the ICG vial to mix the solution. Ensure all ICG powder is
dissolved in the solution.
6. Inspect the reconstituted solution for precipitation. If a precipitate is
present, discard the solution and prepare a new vial as described above.
ICG Dosage
The typical dose administered is 2.5mg.
The total dose should be less than 2 mg/kg per patient.