28
Error Messages
In the event of a light source error, an error message will display on the
touchscreen. See the table below for details.
If the light source displays any error message, reboot the console. If the
error recurs, it is recommended that the device be returned for service.
Code Definition
Details
E-1
Over
temperature
(safety shutoff)
Output in White Light mode is disabled due to
excessive heat in the console. All functionality is
disabled.
E-2
Failure to color
balance
The light source will not actively balance the light
output color.
E-3
Power supply
fan failure
The light source continues operating but generates
excessive heat which may result in additional
damage and/or compromised functionality.
E-4
Heat sink fan
failure
The light source continues operating but generates
excessive heat which may result in additional
damage and/or compromised functionality.
E-5
Overcurrent
LED failure
The light output color may drastically change, and
the light source will not actively balance the light
output color.
E-6
Software
version
mismatch
All light source functionality is disabled. Return to
Stryker for service.
E-7
ENV power
balance failure
The light source will not actively balance the
backlight color until the console is rebooted.
E-8
IRIS power
balance failure
IRIS mode is disabled.
E-9
ENV laser error ENV mode is disabled.
E-10 System
initiation error
All light source functionality is disabled. Return to
Stryker for service.