1-54
280C-009-002 REV B
English
6 . Using diagonal pliers, cut the two cable ties holding the defective cell cable to the frame immediately after the cell .
Repeat for the adjacent cell .
7 . Lower the bed until the two Allen screws holding the defective load cells to the frame are easily accessible .
8 . Unplug the power cord from the wall receptacle .
9 . Using a 1/4” Allen wrench and a 1/2” combination wrench, remove the two nuts and two Allen screws (B) holding
the defective load cell to the frame .
10 . Cut the cable ties holding the defective load cell cable the frame . Unplug the defective load cell cable from the
scale/bed exit board . Remove the defective load cell .
11 . Inspect the elastomer sleeves (C) to ensure that the inner circular hole is not worn . Replace if necessary . Note the
orientation of the elastomer sleeves when replacing (see detail at right) .
12 . Reverse the above steps to install the new load cell .
13 . Calibrate the scale before returning the bed to service . Refer to the “Scale
Calibration” procedure,
.
LOAD CELL REPLACEMENT (OPTIONAL) (CONTINUED)
Support
Point
Support
Point
Service Information
Proper elastomer
sleeve orientation.
Содержание FL28C
Страница 2: ......
Страница 83: ...www stryker com 280C 009 002 REV B 1 81 Return To Table of Contents English Notes ...
Страница 94: ...1 92 280C 009 002 REV B www stryker com Return To Table of Contents English Notes ...
Страница 111: ...www stryker com 280C 009 002 REV B 1 109 Return To Table of Contents English Notes ...
Страница 170: ......
Страница 251: ...www stryker com 280C 009 002 REV B 2 81 Retour à la table des matières Français Commentaires ...
Страница 262: ...2 92 280C 009 002 REV B www stryker com Retour à la table des matières Français Commentaires ...
Страница 279: ...www stryker com 280C 009 002 REV B 2 109 Retour à la table des matières Français Commentaires ...
Страница 336: ......
Страница 337: ......