During reinstall, align the V
Ve
ellccrro
o® patches.
•
During reinstall, align the V
Ve
ellccrro
o® patches.
•
After installation, remove the zipper pull tool or equivalent from the zipper.
•
Cover the zipper with the support surface cover watershed.
13. Verify proper operation before you return the product to service.
P
Pu
um
mp
p a
asssse
em
mb
bllyy rre
ep
plla
acce
em
me
en
ntt
T
To
oo
ollss rre
eq
qu
uiirre
ed
d::
•
Zipper pull tool or equivalent
•
Diagonal pliers
•
T15 Torx driver
P
Prro
occe
ed
du
urre
e::
1. See
2. Disconnect the pump power connector from the main board (connector F).
3. Disconnect the cable from the three wire retainers.
4. Using diagonal pliers, cut the cable ties where the inlet hose connects to the manifold (AY) and the exhaust hose
connects to the resonator (Figure 17). Do not cut the manifold barb.
N
No
otte
e -- During installation, before you cut the end off of the cable tie, use the diagonal pliers to pry the slack out of the
cable tie (one to two clicks).
2973-009-002 Rev AB.0
37
EN
Содержание 297300000000
Страница 2: ......
Страница 13: ...Product serial number Completed by Date 2973 009 002 Rev AB 0 9 EN ...
Страница 52: ...EN 48 2973 009 002 Rev AB 0 ...
Страница 53: ...2973 009 002 Rev AB 0 49 EN ...
Страница 54: ...EN 50 2973 009 002 Rev AB 0 ...
Страница 55: ...2973 009 002 Rev AB 0 51 EN ...
Страница 58: ...F Fo oo ot tb bo ox x a as ss se em mb blly y 297300220008 Rev AD Reference only EN 54 2973 009 002 Rev AB 0 ...
Страница 59: ...2973 009 002 Rev AB 0 55 EN ...
Страница 60: ...EN 56 2973 009 002 Rev AB 0 ...
Страница 61: ...2973 009 002 Rev AB 0 57 EN ...
Страница 62: ...EN 58 2973 009 002 Rev AB 0 ...
Страница 63: ...2973 009 002 Rev AB 0 59 EN ...
Страница 64: ...EN 60 2973 009 002 Rev AB 0 ...
Страница 65: ...2973 009 002 Rev AB 0 61 EN ...
Страница 76: ...Stryker Medical 3800 E Centre Avenue Portage MI 49002 USA 2973 009 002 Rev AB 0 WCR AB 3 2021 06 ...