F
F
Fiig
gu
urre
e 9
9 –– G
Ge
ell ccrriib
b
13. Using a T15 Torx driver, remove the nine ground screws (B) and the nine star washers (D) (Figure 10). Save the ground
screws and the star washers.
N
No
otte
e
•
During reinstall, tighten the screws equally and do not over tighten. Torque item B to 22 ± 4 in-lb.
•
During reinstall, check the LAL fan cable (G) and grommet (H) are in place before replacing the foot box cover
•
After installation, remove the zipper pull tool or equivalent from the zipper.
14. Remove the foot box cover (AD) (Figure 10). Save the foot box cover.