5. Turn off the battery disconnect switch to turn the bed off.
6. See
7. Using diagonal pliers, cut the cable tie (AY) that secures the power/data cable (BC) to the foot box. Discard the cable tie.
N
No
otte
e -- During installation, tighten the cable tie (AY) to 25 - 40 lb.
F
Fiig
gu
urre
e 1
15
5 –– P
Po
ow
we
err//d
da
atta
a cca
ab
blle
e
8. Using a T15 Torx driver, remove the screw (B) from the cable clamp (BL) (Figure 15). Save the screw.
9. Remove and save the cable clamp (B) (Figure 15).
10. Unlock both connectors (BD, BK) and the ends of the power/data cable (BC) to disconnect the power/data cable.
Discard the power/data cable.
11. Reverse steps to reinstall. Orient the cable clamp (BL) as shown (Figure 15).
N
No
otte
e -- During reinstall, torque the screw (B) 22 ± 4 in-lb.
12. Route the power/data cable (BC) through the bottom cover (Figure 16).
EN
36
2973-009-002 Rev AB.0
Содержание 297300000000
Страница 2: ......
Страница 13: ...Product serial number Completed by Date 2973 009 002 Rev AB 0 9 EN ...
Страница 52: ...EN 48 2973 009 002 Rev AB 0 ...
Страница 53: ...2973 009 002 Rev AB 0 49 EN ...
Страница 54: ...EN 50 2973 009 002 Rev AB 0 ...
Страница 55: ...2973 009 002 Rev AB 0 51 EN ...
Страница 58: ...F Fo oo ot tb bo ox x a as ss se em mb blly y 297300220008 Rev AD Reference only EN 54 2973 009 002 Rev AB 0 ...
Страница 59: ...2973 009 002 Rev AB 0 55 EN ...
Страница 60: ...EN 56 2973 009 002 Rev AB 0 ...
Страница 61: ...2973 009 002 Rev AB 0 57 EN ...
Страница 62: ...EN 58 2973 009 002 Rev AB 0 ...
Страница 63: ...2973 009 002 Rev AB 0 59 EN ...
Страница 64: ...EN 60 2973 009 002 Rev AB 0 ...
Страница 65: ...2973 009 002 Rev AB 0 61 EN ...
Страница 76: ...Stryker Medical 3800 E Centre Avenue Portage MI 49002 USA 2973 009 002 Rev AB 0 WCR AB 3 2021 06 ...