Neptune S Rover (REF 0711-001-000)
P
Po
ow
we
err R
Re
eq
qu
uiirre
em
me
en
nttss::
120 V
(alternating current AC), 60 Hertz (Hz), X.0 Amps (A), single
phase; X0 V
(direct current DC), X A during docking procedure; rover
receives power from docker REF 0702-014-000
D
Diim
me
en
nssiio
on
nss::
XX mm [X.X inch] height, XX mm [X.X inch] width, XXX mm [X.X inch] length
M
Ma
assss::
85 kg [187 lb.] — collection canister full; 106 kg [233 lb] — fully loaded, includes
maximum supported load of 21 kg [46 lb] on the work surface
N
No
otte
e -- A Stryker SafeAir Combi or Compact Console may be attached to
the work surface of the rover using the Neptune S Console Bracket. See
M
Ma
atte
erriia
all::
REACH, RoHS materials?
Reticulated foam that prevents frothing: PVC Coated Polyester 10 PPI.
M
Mo
od
de
e o
off O
Op
pe
erra
attiio
on
n::
Continuous
S
So
ou
un
nd
d P
Prre
essssu
urre
e::
Medium Priority Alert, X - X dB (not adjustable, see IEC 60601-1-8: 200X
(Second Edition) + AM1:20XX)
A
Ad
djju
usstta
ab
blle
e S
Su
uccttiio
on
n L
Liim
miitt::
50 to 520 mm-Hg; measured with all ports closed
N
No
otte
e -- Suction limits remain adjustable in increments of 5 mm-Hg for
settings between 50 to 120mm-Hg. Suction limits increment at 20 mm-Hg for
settings over 120 mm-Hg.
V
Va
accu
uu
um
m M
Me
ea
assu
urre
em
me
en
ntt A
Accccu
urra
accyy::
± 5% of full scale (± 26 mm-Hg, ± 3.47 kPa)
include kPa accuracy?
S
Su
uccttiio
on
n L
Liim
miitt A
Accccu
urra
accyy::
< 26 mm-Hg (3.47 kPa) or 10% of setting
V
Vo
ollu
um
me
e::
8-liter capacity (8-liter canister)
N
No
otte
e -- The canister has increment marks of 10 mL.
V
Vo
ollu
um
me
e O
Ovve
errffiillll P
Prro
otte
eccttiio
on
n::
Software and Mechanical Float? Do we need more detail here?
V
Vo
ollu
um
me
e M
Me
ea
assu
urre
em
me
en
ntt A
Accccu
urra
accyy::
8-liter canister, ± 120 mL
N
No
otte
e -- Volume measurement
accuracy specified does not account
for fluid evaporation or an inclined
plane of operation that exceeds the
specified range.
IIn
ncclliin
ne
ed
d P
Plla
an
ne
e o
off O
Op
pe
erra
attiio
on
n::
± 2.5 degrees
E
Eq
qu
uiip
pm
me
en
ntt T
Tyyp
pe
e::
Type CF Applied Part
E
Eq
qu
uiip
pm
me
en
ntt C
Clla
assssiiffiicca
attiio
on
n::
Class I Medical Electrical (ME) Equipment
IIn
ng
grre
essss P
Prro
otte
eccttiio
on
n ((IIP
P))::
IPX0
EN
66
700000969828 Rev-AA
DRAFT
NOT
FOR
SALE