4. Inspect the rover. See
5. Refill the manifold holder with new, unused manifolds, as required. See
(page 12) for ordering information.
6. If the rover will be used again, use the rover handle to push and relocate the rover. See
(page 36). If
the rover will no longer be used, use the rover handle to push and relocate the rover to a storage area. See
5
5..2
2.. T
To
o IIn
nssp
pe
ecctt tth
he
e E
Eq
qu
uiip
pm
me
en
ntt
N
No
otte
e -- Only biomedical equipment technicians trained and experienced in the maintenance of this reusable medical
device should inspect and maintain this equipment.
W
WA
AR
RN
NIIN
NG
G -- Upon initial receipt and before each use, inspect each component for damage. Do not use the product if
damage is apparent.
W
WA
AR
RN
NIIN
NG
G -- Always comply with the inspection interval to ensure the safe and effective use of the equipment.
W
WA
AR
RN
NIIN
NG
G -- Do not disassemble, modify, or repair this product without the authorization of the manufacturer. Contact
Stryker for service.
N
No
otte
e -- For service, contact your Stryker sales representative or call service. See
(page 4). Outside
the US, contact your nearest Stryker subsidiary.
N
No
otte
e -- Maintenance documentation for this equipment is available upon request to Stryker-authorized service
personnel only.
N
No
otte
e -- Routine and careful inspection of the equipment is the best method for determining the service life of the
equipment. See Inspection Criteria and Actions.
T
Ta
ab
blle
e 2
29
9 –– IIn
nssp
pe
eccttiio
on
n C
Crriitte
erriia
a a
an
nd
d A
Accttiio
on
nss
IIn
ntte
errvva
all
C
Crriitte
erriia
a
A
Accttiio
on
n
Before initial use
Inspect the equipment for damage or
missing components.
If damage is apparent, replace the
equipment. See
section.
Make sure the rover and docker operate
properly.
See
(page 54) and
Before each use and after each
cleaning and disinfection.
Inspect the equipment for damage or
missing components.
If damage is apparent, replace the
equipment. See
Inspect the equipment for corrosion,
discoloration, pitting, cracked materials,
or unacceptable deterioration on any
external surfaces, including labels and
product markings.
Inspect the canister exterior, smoke
evacuator filter cover, and infrared
communication window for cracks or
damage.
Inspect the caster and make sure the
locks function properly.
Inspect the power cord for cuts and the
power cord plug for bent pins.
700000969828 Rev-AA