e -- A SET UP SYSTEM screen will appear () after confirmation of the safety information. A valid manifold must be
inserted into the manifold receptacle to enable suction control.
1. Obtain a new manifold and remove the manifold from its packaging.
2. Insert the new (valid) manifold into the manifold receptacle. Confirm on screen that the manifold is new and unused.
N
No
otte
e -- If the manifold is valid and inserted correctly, a SYSTEM READY indication will occur, a manifold indicator will
illuminate white (NEW) or pink (USED), and suction control will be enabled (START SUCTION button enabled). See
(page 74) for the definitions of a new, valid, used, expired, and invalid manifold.
3. Make sure all the inlet ports of the manifold are closed.
4. Touch the START SUCTION button (Figure 17).
F
Fiig
gu
urre
e 1
17
7 –– T
To
o S
Stta
arrtt a
an
nd
d A
Ad
djju
usstt S
Su
uccttiio
on
n
5. Adjust the manual suction control dial to the maximum suction level.
6. Verify the actual suction value changes as indicated on the main control panel display and reaches a suction level of at
least 494 mm-Hg. If not, see
7. Discard the manifold used for testing purposes.
700000969828 Rev-AA
33
EN