The rover will not dock or an error has occurred during
docking.
The docker power cord is not connected or is loosely
connected. Make sure the docker power cord is connected
securely. See
The docker power switch is in the OFF position. Make sure
the power switch is in the ON position and illuminated.
If power switch is OFF, push the power switch to the ON
position. Wait 60 seconds. Dock the rover.
The docker power switch is in the ON position, but not
illuminated. Make sure facility power is provided to the
hospital-grade power receptacle. If facility power is OFF,
apply facility power to the docker. Wait 60 seconds. Dock
the rover.
The rover is not completely connected to the docker. Push
the rover forward enough to complete a docker-rover
interface connection. See
Communication between the rover and docker is
interrupted because the infrared communication window is
either dirty or obstructed. Remove the dirt or debris from
the rover’s infrared communication window. See
(page 50). Dock the rover.
Remove any obstruction from the docker’s infrared
communication windows. Make sure there are no tubes,
hoses, or towels covering the docker’s infrared
communication windows. Dock the rover. See
The docker is not receiving facility water. Make sure the
water inlet hose is connected correctly and the facility water
supply valve is open. Dock the rover.
The docker requires a power reset. Remove power, then
apply power to the docker. Wait 60 seconds. Dock the
rover.
If the problem persists, the docker may be damaged.
Contact Stryker Neptune Customer Service. See
Prefill fluid remains inside the rover canister after a cycle is
complete.
The presence of prefill is normal. This water and detergent
solution (prefill) initiates the breakdown of fluid waste
collected during the rover’s next use. No action is required.
The docker does not dispense detergent during the
cleaning cycle.
The bottle of detergent is empty. Replace the bottle of
detergent. See
The end of the detergent inlet tube is not immersed in the
detergent. Make sure the detergent inlet tube is routed
through the tube stiffener and the end of the tube reaches
the bottom of the detergent bottle.
The detergent inlet tube is not connected securely to the
detergent inlet port of the docker. Securely connect the
detergent inlet tube to the detergent inlet port of the docker.
EN
60
700000969828 Rev-AA
DRAFT