System
Number
Error
Possible Cause
Trouble Shooting
DOCKING
3.0
Docker
communication
error
- Rover main control board bad
- Rover IR board bad
- Docker IR board bad
- Wiring harness bad
Problem is most likely due to IR
communication failure. Check
IR to make sure it is connected
properly and is not obstructed.
If a component has failed you
will most likely receive a
"Docker not Ready" fault as
well.
3.1
Coupling
Extend
Timeout
- Diverter valve failed
- Extend hall sensor failed
- Physical obstruction in the waste
couplers
- Manually fill water in large
tank to verify diverter is
switching properly (using dock
control)
- Refer to procedure 6.8
Manually Raising and
Lowering Docker Coupler
on
page 6-43 to verify proper hall
sensor operation.
- Check for obstruction in
waste coupler
3.2
Coupling
Retract
Timeout
- Retract hall sensor bad
- Stepper motor failed or mounting
hardware loose
- Verify stepper motor is
mounted properly.
- Replace the retract hall
sensor.
3.3
Offload error -
small tank
- Fluid is not offloading due to clog.
- Something preventing float from
reaching the bottom of tank.
- Small canister calibration board
damaged.
- Level transducer damaged.
- Calibration is incorrect.
- Verify that something is not
blocking the drain valve from
emptying.
- Verify nothing is underneath
the float.
- Recalibrate the canister.
- Replace small tank float
sensor
- Replace small canister
- Replace fluid level
transducer
3.4
Offload error -
large tank
- Calibration is incorrect
- Clog or obstruction in the large
canister
- Offload pump not running
- Recalibrate large canister
- Check for and remove clog
or obstruction
- Check for proper voltage at
offload pump
3.5
Prefill error -
small tank
- Calibration is incorrect;
- Water has been shut off
- Recalibrate
- Turn water on
3.6
Prefill error -
large tank
- Calibration is incorrect;
- Water has been shut off
- Recalibrate
- Turn water on
3.7
Extend /
Retract Sensor
Error
- Water on the sensors.
- Faulty sensors.
- Faulty docker control board
- Faulty wiring harness
- Ensure sensors are not wet.
- Replace sensors.
- Replace docker control
board.
- Check wiring harness.
5-23
Содержание 0702-002-620
Страница 1: ......
Страница 3: ......
Страница 5: ......
Страница 15: ......
Страница 17: ......
Страница 19: ......
Страница 21: ......
Страница 22: ...Chapter 1 Operational Description...
Страница 23: ......
Страница 27: ......
Страница 28: ...Chapter 2 Ultra Rover Disassembly...
Страница 29: ......
Страница 31: ...4 Remove front cover assembly P N 0702 001 070 CH and place aside 2 2...
Страница 37: ...4 Remove front cover assembly P N 0702 001 070 CH and place aside 2 8...
Страница 42: ...4 Remove front cover assembly P N 0702 001 070 CH and place aside 2 13...
Страница 49: ...4 Remove front cover assembly P N 0702 001 070 CH and place aside 2 20...
Страница 81: ...4 Remove front cover assembly P N 0702 001 070 CH and place aside 2 52...
Страница 103: ......
Страница 105: ......
Страница 106: ...Chapter 4 Docker Disassembly...
Страница 107: ......
Страница 133: ......
Страница 134: ...Chapter 5 Troubleshooting...
Страница 135: ......
Страница 167: ......
Страница 168: ...6 33 Chapter 6 Technical Procedures...
Страница 169: ......
Страница 219: ......
Страница 220: ...Chapter 7 Installation...
Страница 221: ......
Страница 224: ...7 3...
Страница 228: ...Chapter 8 Appendices...
Страница 229: ......
Страница 231: ......