6.5.3
The electrolyte usage counter
The electrolyte counter registers the number of times a particular electrolyte has been used. This
makes it easier to know when the electrolyte cartridge must be refilled.
You can check which electrolyte is currently registered and check the usage counter.
1.
In the
Main menu
screen, press
Escape
.
2.
A pop-up window will display information about the electrolyte
being used.
6.5.4
Replacing the electrolyte cartridge
CAUTION
Always request and read the Safety Data Sheet for each electrolyte before you start
working with it.
CAUTION
The operator must be fully instructed in how to handle and use electrolytes with this
machine.
CAUTION
The machine is designed to be used with electrolytes recommended by Struers.
Electrolytes that are not recommended by Struers can be dangerous to the operator
or harm the machine.
If you are prompted to change the electrolyte
If a method requires an electrolyte other than the one currently being used, you will be prompted to
replace the electrolyte cartridge.
You will also be prompted to carry out a cleaning step for flushing out the electrolyte with water.
1.
If you wish to change to a new type of electrolyte, press
Enter
.
2.
Replace the electrolyte cartridge containing the current
electrolyte with a cartridge containing the new electrolyte.
See
Inserting the electrolyte cartridge ►32
.
3.
If you wish to continue using the current electrolyte, press
Esc
.
Changing the electrolyte from the Maintenance screen
From the
Maintenance
screen, you can change the electrolyte and select a step for flushing out the
electrolyte with water.
34
MoviPol-5
6 Operating the device