Error Message
Explanation
Action
3
Fatal error
Machine failed during
Power On Self Testing.
Please reboot the machine.
If the problem persists
please contact Struers
technical support.
Reason: #__ - Unknown
error
Failure of internal
communication on start-up.
Re-start the machine.
If the error remains, contact
Struers Service.
Make a note of the Reason
number.
28
Error
Specimen holder cannot be
moved down.
Down proximity sensor has
not detected bottom
position.
Make sure that there are no
obstacles preventing the
movement of the specimen
holder.
Check the pneumatic
system.
If the error remains, contact
Struers Service.
29
Warning
No air or air pressure too
low!
Pressure of the
compressed air supply is
too low.
Check the compressed air
supply.
30
Error
Pressure regulating error!
Pressure of the
compressed air supply is
too high/low.
Check the compressed air
supply.
Re-start the machine.
If the error remains, contact
Struers Service.
34
Error
Specimen mover plate
cannot be moved down.
Down proximity sensor has
not detected bottom
position.
Make sure that there are no
obstacles preventing the
movement of the specimen
mover plate.
Check the pneumatic
system.
If the error remains, contact
Struers Service.
46
Error
Disc motor RPM's has not
been reached.
Disc motor will not rotate or
cannot reach the set rpm.
Polishing process is
paused.
Re-start the process.
If the error remains, contact
Struers Service.
26