Inserting a flexible specimen holder
CAUTION
To prevent specimens from detaching from the specimen holder, make sure that the
specimen or specimens are covered completely by the flexible specimen holder.
CRUSHING HAZARD
Keep your hands clear of the flexible specimen holder when lowering the specimen
mover.
Procedure
Hint
Make sure that you use sufficient force according to the specimen size and Struers
recommendations.
The Struers Metalog Guide methods are based on a specimen with an area of 7 cm
2
.
Adjust the method according to your specific specimen area.
Procedure
Hint
Make sure that the preparation surface is sufficiently wet before you start the
preparation process.
1.
Press the
Lower/Raise
button to make sure that the specimen mover head is
fully raised.
2.
Press the black button on the specimen mover head.
3.
Insert the flexible specimen holder and rotate it until the three pins are aligned
with the holes in the specimen mover.
4.
Push the flexible specimen holder upwards until it locks into position.
5.
Release the black button on the specimen mover head. Make sure that the
flexible specimen holder is securely fixed.
Using the flexible specimen holder
1.
Place the specimen or specimens on the preparation surface.
2.
Press the
Lower/Raise
button to lower the flexible specimen holder.
3.
Make sure that no specimens protrude from the flexible specimen holder. If they
do, adjust the specimens.
–
Press the
Lower/Raise
button to raise the flexible specimen holder.
–
Adjust the specimens.
4.
Repeat until all specimens are placed correctly.
5.
Start the preparation process.
The preparation process stops automatically when the set preparation time has
expired.
6.
Clean the flexible specimen holder before the next preparation step.