Holder up
Moves the specimen mover head up for cleaning.
–
Holder down
Moves the specimen mover head back into operating
position.
7.4
Monthly
7.4.1
LaboForce-100 - Emptying the water/oil filter
The specimen mover is fitted with a water/oil filter that removes excessive amounts of water and
oil from the compressed air supply.
The filter must be emptied on a regular basis.
Procedure
1.
Locate the release outlet valve at the bottom of LaboForce-
100.
2.
Hold a cloth under the release outlet valve and press the valve
to empty the water/oil filter.
7.5
Spare parts
For specific safety related parts, see the section "Safety Related Parts of the Control System
(SRP/CS)" in the section "Technical data" in this manual.
Technical questions and spare parts
If you have technical questions or when you order spare parts, state serial number and
voltage/frequency. The serial number and the voltage are stated on the type plate of the machine.
For further information, or to check the availability of spare parts, contact Struers Service. Contact
information is available on
7.6
Service and repair
Struers recommends that a regular service check be carried out yearly or after every 1500 hours
of use.
When the machine is started up, the display shows information about total operation time and the
machines service information.
After 1500 hours of operation time, the display will show a message reminding the user that a
service check should be scheduled.
Note
Service must only be performed by a Struers engineer or a qualified technician
Contact Struers Service.
7 Maintenance and service
LaboForce-100
23