5.5
Compressed-air connections
Procedure
Note
The main air valve is not a part of the unit and must be installed and set before the
specimen mover is installed.
1.
Fit the quick coupling on the compressed air hose and secure it with the hose clamp.
2.
Connect the air inlet hose to the quick coupling.
3.
Fit the other end of the air inlet hose into the compressed air inlet on the specimen mover.
Note
The air pressure must be between 6 bar (87 psi) and 9.9 bar (143 psi).
Hint
The specimen mover requires a continuous flow of compressed air through the
regulator valve – a faint hissing sound does not mean that there is an air leak.
5.6
The specimen mover
The specimen mover can be used with specimen mover plates for single specimens or specimen
holders for multiple specimens.
5.6.1
Inserting a specimen holder
Inserting a specimen holder
CAUTION
To prevent specimens from detaching from the specimen holder, make sure that
the specimen or specimens are clamped securely in the specimen holder.
CAUTION
Protective footwear is recommended when you handle heavy specimen holders.
CRUSHING HAZARD
Keep your hands clear of the specimen holder or specimen mover plate when
lowering the specimen mover.
Note
When you work with specimen holders, make sure that the screws clamping the
specimens do not stick out of the specimen holder.
Use different lengths of screws for specimens with different diameters.
5 Installation
LaboForce-100
11