Maintenance, servicing, repairs, and disposal
Instructions for use • AIDA C WD310 • AAB272_EN_V2.0_12-2021_IFU_CE-MDR
92
8 Maintenance, servicing, repairs, and disposal
8.1 Maintaining the product
If they are not described in more detail here, maintenance activities may only be performed by
KARL STORZ or by a company authorized by KARL STORZ.
8.1.1 Maintenance
The following maintenance intervals are recommended:
Interval
Activity
To be performed by
As required
Change fuse
KARL STORZ service techni-
cian or a qualified person
Annually
Safety inspection
KARL STORZ service techni-
cian or a qualified person
Every 4 years
Replace lithium battery
KARL STORZ service techni-
cians
8.2 Safety inspection in accordance with IEC 62353
WARNING
Risk of injury due to product deficiencies!
Patients, users, and third parties may be injured as a result of deficiencies with the product and
accessories.
Shut down the product.
Have the deficiencies repaired by persons authorized by KARL STORZ.
Regardless of the national accident prevention regulations and testing intervals for medical
devices, for this device safety checks must be performed as repeat inspections according to
IEC 62353 and recorded by a qualified electrician at least once a year. Detailed specifications
on the scope and performance of the safety check can be found in the service manual.
8.2.1 Visual inspection
1.
Check the product and accessories for any mechanical damage.
2.
Check labels for readability.
8.2.2 Electric measurements
Limit values for electrical measurements can be found in the current IEC 62353.
1.
Measure the protective ground resistance.
2.
Measure the earth leakage current.
3.
Measure the touch current.
4.
Measure the patient leakage current.
8.2.3 Functional test
1.
Perform a functional test in line with the instructions for use.
2.
Document the results of the safety inspection.
Содержание AIDA C WD310
Страница 1: ...Instructions for use AIDA C WD310 Release 1 1 en...
Страница 113: ...Subsidiaries Instructions for use AIDA C WD310 AAB272_EN_V2 0_12 2021_IFU_CE MDR 113...
Страница 114: ...Subsidiaries Instructions for use AIDA C WD310 AAB272_EN_V2 0_12 2021_IFU_CE MDR 114...
Страница 115: ...Subsidiaries Instructions for use AIDA C WD310 AAB272_EN_V2 0_12 2021_IFU_CE MDR 115...