Installation and commissioning
Instructions for use • AIDA C WD310 • AAB272_EN_V2.0_12-2021_IFU_CE-MDR
28
1.
In the Client AET field, enter the AE title which was assigned to the AIDA C system by the
IT department.
2.
Open the Modality drop-down menu and select a type:
ES: endoscopy
US: ultrasound
OT: other
3.
In consultation with the IT department, select a Char. set.
4.
Leave the LogLevel as the factory setting.
5.
Place a check mark next to Enable Worklist to allow access to a DICOM worklist.
ð
The Import DICOM Worklist button is displayed on the user interface.
6.
Enter the following in the fields:
Server AET: AE title of the DICOM worklist server
Server IP: IP address of the DICOM worklist server
Port: port of the DICOM worklist server
7.
Open the Date Filter drop-down menu and limit the date range displayed when searching
for patient data in the DICOM worklist:
None: date range not limited
Today: only display the patients in today's worklist
One Day: plus or minus 1 day
Two Days: plus or minus 2 days
Three Days: plus or minus 3 days
8.
Leave the Timeout as the factory setting unless it needs to be changed.
9.
Tap the Echo button to test the connection to the DICOM server.
10. In the event of connection problems, contact the IT department.
Содержание AIDA C WD310
Страница 1: ...Instructions for use AIDA C WD310 Release 1 1 en...
Страница 113: ...Subsidiaries Instructions for use AIDA C WD310 AAB272_EN_V2 0_12 2021_IFU_CE MDR 113...
Страница 114: ...Subsidiaries Instructions for use AIDA C WD310 AAB272_EN_V2 0_12 2021_IFU_CE MDR 114...
Страница 115: ...Subsidiaries Instructions for use AIDA C WD310 AAB272_EN_V2 0_12 2021_IFU_CE MDR 115...