Table 206
Recommended separation distances between portable and mobile RF communications equipment and the
XENON 300 model 20
1331 20-1
The XENON 300 model
20
1331 20-1 is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled.
The customer or user of the unit can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF
communications equipment (transmitters) and the unit as recommended below, according to the maximum output power of the communications
equipment.
Separation distance d
[
m
]
according to frequency of transmitter
Rated maximum
output power of
transmitter
[W]
150 kHz to < 80 MHz
P
d
⎥⎦
⎤
⎢⎣
⎡
=
3
5
.
3
80 MHz to < 800 MHz
P
d
⎥⎦
⎤
⎢⎣
⎡
=
3
5
.
3
800 MHz to 2.5 GHz
P
d
⎥⎦
⎤
⎢⎣
⎡
=
3
7
0.01
0.12
0.12
0.23
0.1
0.38
0.38
0.73
1
1.2
1.2
2.3
10
3.8
3.8
7.3
100
12
12
23
For transmitters rated at a maximum output power not listed above, the recommended separation distance
d
in meters (m) can be estimated using the
equation applicable to the frequency of the transmitter, where
P
is the maximum output power rating of the transmitter in watts (W) according to the
transmitter manufacturer.
Note: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects