23
Stille imagiQ2
TM
user manual
Care
n
C
are
Maintenance
Periodical preventive maintenance (PM)
The imagiQ2 requires a maintenance period of 1time
every year or 2000 hours of service whichever comes
first according to the service schedule in this manual.
(2000 hours equals to 8 hours of service 5 days / week
for 1 year)
If the service is not correctly documented the warranty
is void.
Stille AB can not guarantee the product’s safety if Periodi-
cal preventive maintenance (PM) has been neglected.
The device should be cleaned routinely, as described
below. Additional preventive
maintenance is required according to the Service manual
Cleaning
The table should be thoroughly cleaned, after each
procedure (follow hospital cleaning procedure). Clean
the device as follows, see special reccomendations in the
service manual regarding cleaning and disinfection of the
mattress.
Do not use steam cleaning.
Not intended to be sterilized.
Inspection of contamination on the device shall be
done before each use.
The mattress shall be inspected on the regular
service interval. Scratches or holes are not accepta-
ble.
Unplug the power connector from its power source
before moving and/ or cleaning. Failure to do so could
result in equipment damage.
When cleaning the imagiQ2™ table and its acces-
sories, use a lightly alkaline universal cleaning agent
(soap water).
The cleaning agent should contain phosphates and
tensides as active ingredients.
When surfaces are heavily soiled, concentrated
cleaners can be used. In this case the table should be
wiped off with clean water afterwards.
Disinfect the metal surfaces of the operating table
with disinfectants that do not contain chlorine or
compounds that give off chlorine on decomposition.
Do not use steam cleaning.
The
imagiQ2™ table should only be cleaned and disin-
fected by hand..
Cleaning and care of the polyurethane coated mattress
General Guidance
Attention must be paid to the properties of any other
materials, (eg washing agents and washing instruc-
tions).
Some surface wrinkling may result from cleaning
procedures. This has no adverse effect on the fabric’s
properties.
Do not use washing machines
Abrasive cleaning agents should not be used.
Washing and Disinfection
Superficial dirt on the coating may be removed by
wiping with a soft cloth moistened with water containing
a neutral detergent. More persistent contamination may
be treated by wiping with alcohols or turpentine substi-
tute, followed by hot water and detergent.
Routine cleaning and disinfection in situ may be car-
ried out on the coating with hand hot water and a neutral
detergent or with a sodium hypochlorite solution (0.1%
or 1000 parts per million available chlorine).
The material is compatible with the 10,000ppm avai-
lable chlorine in solution required for the decontamination
of blood spills. After the 2 minute required dwell, excess
solution must be removed and the surface thoroughly
rinsed and dried prior to reuse or storage.
Proprietary disinfectants may be used provided
manufacturer’s instructions are followed.
All cleaning agents, and disinfectants, must be
thoroughly rinsed off and the item dried before storage.
Failure to do this may result in the accumulation of rea-
gent that could damage the polyurethane coating, react
with the bed frame, or negate the biocompatibility results
of the fabric.
Drying
It is essential that articles be thoroughly dried after all
cleaning procedures and before storage.
Drying may be effected at temperatures up to 130
°C (266°F).
Storage
Store in a cool, dry area. Avoid excessive pressure and
contact with non-colourfast materials