22
Stille imagiQ2
TM
user manual
Care
n
C
are
Charging
The imagiQ™ table has an integrated, fully automatic
charging unit. Regular charging gives high operational
reliability and a long life for the batteries.
The mains voltage is 100 - 230 V AC. An indicator in
the fuse-holder integrated
in the mains inlet shows the preset voltage setting.
The secondary voltage is 24 V DC.
The ImagiQ2™ table is approved for charging in the
operating theatre with a patient on the table.
The batteries are NOT being charged when the
Quick-Stop is off, (Note: the blue LED on the hand
control will not light up if the Quick-Stop is off).
If the batteries are not charged correctly they might
become damaged.
See page 17 for additional battery information
CARE and MAINTENANCE
1.
Constant blue light indicates the charging is in progress.
The table is connected to mains.
2.
Constant blue and green light indicates
the batteries are
fully charged. Maintenance charging in progress
1.
2.
Charging instructions
1. Park the table.
2. Connect mains cord and check that the blue LED on
the hand control lights up.
3. Charging is complete when both blue and green LED’s
are lit constantly.
Note 1: The mainswitch must be turned ON
Note 2: A damaged mains cord must be replaced imme-
diately. Always use imagiQ2™ table original mains
cord.
Note 3: Unplug the mains cord before unparking the table.