17
Stille imagiQ2
TM
user manual
Handling
n
Saf
et
y Inf
or
mation
15. Battery and charging indicators
The device is dual powered, either battery powered when the AC cable is disconnected, or directly powe-
red when AC cable is connected to the device, the batteries are charging when the device is connected
to AC power. The battery indicators are constantly measuring the status of the batteries, during load/
movement the battery indicators might shift from one status to another, but when the load is released
the battery indicator will shift back after a short while, this is normal behaviour.
15.1
Flashing Green light
70-100 % level of charge. Max battery life-time is achieved if charged regularly at this level.
15.2 Flashing Blue light
50 % level of discharge. Batteries should be charged as soon as session is over.
15.3 Rapidly flashing Blue light
90 % level of discharge. Batteries must be charged as soon as possible. If charging process can’t be initiated the device must
be turned off using the OFF button on the hand control. When discharged to this level, the batteries must be charged until full
without interruption, this can take between 12 to 48 hours. If charging is interrupted, batteries may become damaged.
15.4 Constant Blue light
Charging is in progress. The table is connected to mains.
15.5 Constant
Blue and Green light
Batteries fully charged. Maintenance charging in progress
Note:
Blue or Green flashing signals are given only when the system is ON. Blue flashing signal for is given regardless of
system ON or OFF.
Sounds
The imagiQ2 uses different sounds to alert the user of the following conditions:
1. Repeated beep, auto shut off.
If no buttons are pressed for two hours, the table is switched off automatically. The (ON) lamp on the hand
control flashes and a repeated beep gives a ten minute warning before automatic switch-off. Press any button to
prevent automatic switch-off. If no movement is required, press the (ON) butto
n
2. Short repeated beeps, trying to unpark loaded table.
If table is uncentered and then is unparked with >100kg load on the table top the table will sound short repeated beeps and
flash lock unlock on the hand control for ~5 seconds. Place table in safe position (page11)
3. Constant beep, unparked table loaded.
Unparked uncentered table is loaded with ~>100kg a constant beep is heard. Centre the table top laterally, longitudinally and
lower the height in order to clear the alert.
4. Constant beep, over load.
If the table is unparked and the column is all the way up the table will signal overload even without load. Lower the column to
clear the alert.
5. Repeated beebs, battery low.
If the batterys are critically low the table will beep for 1 miute and then turn the table off, if the table is turned on again the
table will start to beep and then turn off again.
6. Short beeps during 1minute then the table turns off.
1. Motor FET fail: 20x5ms beeps
2. X-Brake system fail: 10x10mss beeps
3. Wheel communication fail: 4x250msbeeps
4. SoftwareFailure: 2x500ms beeps
The above trouble messages can be overridden if the ON button is pressed for 10 seconds when the table is beeping. Do not
override the trouble message except in case of emergency or to finalize an ongoing procedure.
7. Continous beep every 30 second.
One of the above failstates have been detected and the table have manually been put into failstate mode, all
safety functions have been disconnected. WARNING! the table may not be transported with a patient on the
table, make sure to lock the parking cirquit all the time when the table is used with a patient on it.
Содержание imagiQ2
Страница 1: ...Stille imagiQ2 User manual Surgical Perfection For life ...
Страница 26: ...NOTES ...
Страница 27: ...NOTES ...