4 Start-up
ASTODIA Instructions for Use
19
4.5 Cleaning and disinfecting
NOTICE
Actions to avoid damaging the diaphanoscope:
- Do not immerse the control unit and/or the connector for the connection cable
in liquid.
- Do not disinfect the system using steam (e.g., in autoclaves), hot air, or
thermo-chemical cleaning solutions.
Clean and disinfect the control unit and the hand piece, if necessary, or when it is
polluted as follows:
1. Clean all surfaces using a soft cloth/cotton swab and a mild soap solution.
2. Disinfect the control unit and the hand piece using a recommended disinfect-
ant.
3. Only disinfect the control unit and the hand piece with a disinfectant on the
basis of alcohol or with one of the recommended disinfectants.
List of approved, ready-to-use disinfectants:
Disinfectant
Manufacturer
Aniosurf ND Premium
Laboratoires ANIOS
Bacillol Plus
BODE Chemie GmbH
BIGUAMED PERFEKT N
Desomed-Dr.Trippen GmbH
Incidin Plus
®
Ecolab Healthcare
Incidin Rapid
Ecolab Healthcare
Meliseptol Foam pure
B. Braun Melsungen AG
Meliseptol Wipes sensitive
B. Braun Melsungen AG
Mikrobac forte
Paul Hartmann AG
Mikrozid AF liquid
Schülke & Mayr GmbH
Super Sani-Cloth
®
Ecolab Healthcare
Terralin protect
Schülke & Mayr
Follow the specific instructions for use.
NOTICE
The operator should not use any cleaning or decontamination methods other
than those recommended by the manufacturer.