3 Installation
ASTODIA Instructions for Use
17
3 Installation
3.1 Initial start-up
Before you put the device into operation, charge the integrated batteries. To do
this, use the belonging charger which is to plug into the interface of the control
unit.
WARNING
Risk of injury!
To prevent the risk of an electric shock, strictly use a charger approved by the
manufacturer.
Charging is complete when the LED on the charger lights up green or alternately
with a short flash of the yellow LED and the green LED.
Prior to first use, perform the following inspections:
Visual inspection
(see Section 6.2.1 Visual Inspection)
4 Start-up
Read each section before you use ASTODIA.
4.1 Preparation for use
To apply the ASTODIA connect the hand piece with the interface of the control
unit.
Make sure that the plug of the connection cable for the
ASTODIA handpiece is fully plugged in. If the plug is not
completely plugged in, the function of the handpiece is lim-
ited; e.g. only the yellow LED lights up.
4.2 Starting
CAUTION
Risk of injury!
•
In case ASTODIA starts directly after switch on in the highest light level and a
change of the highest level is not possible - then do not use ASTODIA!
•
Due to its high light intensity this device may not be used in close vicinity to
the patient's eyes.
•
ASTODIA may not be used several times on a fixed point on the surface of
the skin.
•
Once ASTODIA has been switched on, the red and yellow LEDs of the hand
piece each flash three times (functional testing). ASTODIA may only be used
after the LEDs have flashed. If the LEDs do not flash as described, ASTODIA
must be serviced before being used again.