4-15
Operating Instructions
Operator Manual
122994-513
4.8 Stop Cycle
Operation
NOTE: Washer operation can be halted at any time by pressing
STOP/RESET
touch pad. Pressing
STOP/RESET
touch pad once
halts cycle and printout message says "STOP PRESSED!". Pressing
STOP/RESET
touch pad twice aborts cycle and printout message is
"CYCLE ABORTED!". To resume cycle, press
CYCLE/START
touch
pad. In an emergency, if it is necessary to open door after pressing
STOP/RESET
touch pad, wait for water flow to stop and then open
door. To resume cycle, close door and press
CYCLE/START
touch
pad.
1.
Press
STOP/RESET
touch pad to immediately halt operation of
cycle in progress. Display screen indicates
STOP/RESET
touch
pad was pressed:
alternating with:
and printer records:
WASH1 2:51:32P
RINSE1 2:54:48P
*STOP PRESSED
2:55:00P
2.
Press
CYCLE/START
touch pad to resume cycle operation.
Treatment resumes at point where it was interrupted:
alternating with:
and printer records:
*CYCLE RESUME 2:55:22P
RINSE 2 2:56:07P
STOP WAS PRESSED
PRESS START TO RESUM
LIGHT
HOT
RINSE1
LIGHT
HOT
RINSE1
LIGHT
HOT
DRAINING
TIME = 00:45
STOP
RESET
CYCLE
START
WARNING – BURN HAZARD:
• Except for an emergency, do
not open door when cycle is
in progress. In an emer-
gency, first stop cycle by
pressing STOP/RESET touch
pad and wait for water flow to
stop. Wear appropriate Per-
sonal Protective Equipment
(PPE) whenever reaching
into chamber.
• After pressing STOP/RESET
touch pad, wait until water
flow stops before slowly
opening door. Hot water/
steam may be sprayed
through door opening if
door is opened too soon.
Содержание RELIANCE 400
Страница 6: ...iv 122994 513 Operator Manual Introduction Life Sciences Capital Equipment ...
Страница 10: ...viii 122994 513 Operator Manual Table of Contents Life Sciences Capital Equipment ...
Страница 12: ...x 122994 513 Operator Manual List of Illustrations Life Sciences Capital Equipment ...
Страница 18: ...1 6 122994 513 Operator Manual Safety Precautions Life Sciences Capital Equipment ...
Страница 24: ...2 6 122994 513 Operator Manual Installation Verification Life Sciences Capital Equipment ...
Страница 30: ...3 6 122994 513 Operator Manual Component Identification Life Sciences Capital Equipment ...
Страница 52: ...4 22 122994 513 Operator Manual Operating Instructions Life Sciences Capital Equipment ...
Страница 105: ......
Страница 106: ...STERIS Corporation 5960 Heisley Road Mentor OH 44060 1834 USA 440 354 2600 800 JIT 4 USE 800 548 4873 ...