Safety Precautions
Operator Manual
122994-513
CAUTION – POSSIBLE EQUIPMENT DAMAGE:
1
Vaseline is a registered trademark of Cheseborough Pond's Incorporated.
Always position each manifold and/or bottom rotary spray over a manifold connector before operating
unit. If manifolds and/or bottom rotary sprays are not positioned correctly, damage will result and unit
will be unable to effectively wash load.
Always position each accessory header over a manifold connector before operating unit. If accessory
headers are not positioned correctly, damage may result and unit will be unable to effectively wash load.
Always select a cycle appropriate for the items being processed. Failure to do so may result in product
damage.
Always use a silicone lubricant to lubricate squeeze tubes. Petroleum-based lubricants, such as Vase-
line
®
1
or grease, will cause squeeze tubes to melt.
Once chamber door has been lowered, ensure cable is correctly aligned and routed on all six cable pul-
leys.
Use nonabrasive cleaners when cleaning unit. Follow directions on containers and rub in a back-and-
forth motion (in same direction as surface grain). Abrasive cleaners will damage stainless steel.
Cleaners rubbed in a circular motion or applied with a wire brush or steel wool on door and chamber
assemblies can be harmful to stainless steel. Do not use these cleaners on painted surfaces.
When choosing a detergent, select one with a low-chloride content. Detergents with a high-chloride
content must not be used, as such detergents may harm stainless steel.
Содержание RELIANCE 400
Страница 6: ...iv 122994 513 Operator Manual Introduction Life Sciences Capital Equipment ...
Страница 10: ...viii 122994 513 Operator Manual Table of Contents Life Sciences Capital Equipment ...
Страница 12: ...x 122994 513 Operator Manual List of Illustrations Life Sciences Capital Equipment ...
Страница 18: ...1 6 122994 513 Operator Manual Safety Precautions Life Sciences Capital Equipment ...
Страница 24: ...2 6 122994 513 Operator Manual Installation Verification Life Sciences Capital Equipment ...
Страница 30: ...3 6 122994 513 Operator Manual Component Identification Life Sciences Capital Equipment ...
Страница 52: ...4 22 122994 513 Operator Manual Operating Instructions Life Sciences Capital Equipment ...
Страница 105: ......
Страница 106: ...STERIS Corporation 5960 Heisley Road Mentor OH 44060 1834 USA 440 354 2600 800 JIT 4 USE 800 548 4873 ...