7-9
Troubleshooting
Operator Manual
122994-513
18.
If washer is equipped with power
door:
load side door remains open after
DOOR CLOSE
touch pad was pressed.
Alarm sounds and display shows:
ALARM: LOAD DOOR
DID NOT CLOSE
Printout message:
ALARM: LOAD DOOR HH:MM:SS
DID NOT CLOSE
1.
Air supply valves are closed or air pressure is below specifi-
cations. Press
ALARM REPLY
touch pad to silence alarm
buzzer.
2.
Open building and washer air supply valves and press
DOOR CLOSE
touch pad on load side control panel to
close door.
3.
If condition reoccurs, call STERIS.*
19.
If washer is equipped with power dou-
ble doors:
unload side door was
opened during cycle processing. Alarm
sounds and display shows:
ALARM: UNLOAD DOOR
OPEN DURING PROCESS
Printout message:
ALARM: UNLOAD DOOR HH:MM:SS
OPEN DURING PROCESS
1.
Press
ALARM REPLY
touch pad to silence alarm buzzer.
2.
Close unload side door and press
CYCLE/START
touch
pad to resume operation.
3.
If condition reoccurs, call STERIS.*
20.
If washer is equipped with power dou-
ble doors:
unload side door remains
closed after
DOOR OPEN
touch was
pressed. Alarm sounds and display
shows:
ALARM: UNLOAD DOOR
DID NOT OPEN
Printout message:
ALARM: UNLOAD DOOR HH:MM:SS
DID NOT OPEN
1.
Air supply valves are closed or air pressure is below specifi-
cations. Press
ALARM REPLY
touch pad to silence alarm
buzzer.
2.
Open building and washer air supply valves and press
DOOR OPEN
touch pad on unload side control panel to
open door.
3.
If condition reoccurs, call STERIS.*
Table 7-2. Troubleshooting Chart – Alarm Screen and/or Printout
PROBLEM
POSSIBLE CAUSE AND CORRECTION
*Service charges may be incurred. Consult your warranty for details.
Содержание RELIANCE 400
Страница 6: ...iv 122994 513 Operator Manual Introduction Life Sciences Capital Equipment ...
Страница 10: ...viii 122994 513 Operator Manual Table of Contents Life Sciences Capital Equipment ...
Страница 12: ...x 122994 513 Operator Manual List of Illustrations Life Sciences Capital Equipment ...
Страница 18: ...1 6 122994 513 Operator Manual Safety Precautions Life Sciences Capital Equipment ...
Страница 24: ...2 6 122994 513 Operator Manual Installation Verification Life Sciences Capital Equipment ...
Страница 30: ...3 6 122994 513 Operator Manual Component Identification Life Sciences Capital Equipment ...
Страница 52: ...4 22 122994 513 Operator Manual Operating Instructions Life Sciences Capital Equipment ...
Страница 105: ......
Страница 106: ...STERIS Corporation 5960 Heisley Road Mentor OH 44060 1834 USA 440 354 2600 800 JIT 4 USE 800 548 4873 ...