Operator Manual
Troubleshooting
8.
If washer is equipped with heated pure
water rinse option, optional pure water
tank RTD is detecting out-of-range tem-
peratures. Alarm sounds and display
shows:
ALARM: PW TANK RTD
DEFECTIVE
Printout message:
ALARM: PW TANK RTD HH:MM:SS
DEFECTIVE
1.
RTD connector unplugged. Press
ALARM REPLY
touch
pad to silence alarm buzzer.
2.
Verify RTD connection and press
CYCLE/START
touch
pad to resume operation.
3.
If condition reoccurs, call STERIS.*
9.
Purified water is taking too long to fill.
Alarm sounds and display shows:
ALARM: PURE WATER
TOO LONG IN FILL
Printout message:
ALARM: PURE WATER HH:MM:SS
TOO LONG IN FILL
1.
Pure water supply valves closed. Press
ALARM REPLY
touch pad to silence alarm buzzer.
2.
Open building and washer pure water supply valves and
press
CYCLE/START
touch pad to resume operation.
3.
If condition reoccurs, call STERIS.*
10. If washer is equipped with heated pure
water rinse option, optional pure water
tank RTD is detecting out-of-range tem-
peratures. Alarm sounds and display
shows:
ALARM: PURE WATER
TEMP TOO HIGH
Printout message:
ALARM: PURE WATER HH:MM:SS
TEMP TOO HIGH
Press
ALARM REPLY
touch pad to silence alarm buzzer
and call STERIS.*
Table 7-2. Troubleshooting Chart – Alarm Screen and/or Printout
PROBLEM
POSSIBLE CAUSE AND CORRECTION
*Service charges may be incurred. Consult your warranty for details.
Содержание RELIANCE 400
Страница 6: ...iv 122994 513 Operator Manual Introduction Life Sciences Capital Equipment ...
Страница 10: ...viii 122994 513 Operator Manual Table of Contents Life Sciences Capital Equipment ...
Страница 12: ...x 122994 513 Operator Manual List of Illustrations Life Sciences Capital Equipment ...
Страница 18: ...1 6 122994 513 Operator Manual Safety Precautions Life Sciences Capital Equipment ...
Страница 24: ...2 6 122994 513 Operator Manual Installation Verification Life Sciences Capital Equipment ...
Страница 30: ...3 6 122994 513 Operator Manual Component Identification Life Sciences Capital Equipment ...
Страница 52: ...4 22 122994 513 Operator Manual Operating Instructions Life Sciences Capital Equipment ...
Страница 105: ......
Страница 106: ...STERIS Corporation 5960 Heisley Road Mentor OH 44060 1834 USA 440 354 2600 800 JIT 4 USE 800 548 4873 ...