Reactivating the water softener
Position the salt container as
follows:
1.
Remove the lower basket from the
machine.
2.
Unscrew the plastic cap in the base
of the wash cabinet.
3.
Place the salt container on the reacti-
vation socket and screw firmly into
place.
4.
Close the door.
5.
Turn on the washer/disinfector.
6.
Press the REACTIVATION touch pad.
7.
Press the START touch pad. The
reactivation program proceeds
automatically.
8.
When the REACTIVATION light at
the left hand side of the control
panel and the START indicator light
have gone out, turn off the
washer/disinfector.
9.
Remove salt container slowly,
allowing pressure to be released.
Do not open by force, If you cannot
unscrew it by hand, contact STERIS
Engineering Service.
10.
Screw the plastic lid back on the
socket.
11.
Reposition the lower basket.
12.
Rinse the salt container and filter
insert thoroughly.
IMPORTANT
If, after REACTIVATION cycle, salt is
still found in the salt container, water
pressure may be fluctuating or is below
35 psi (2.5 bar). If this is the case, run
the REACTIVATION cycle again to
wash all water softener from the salt
container.
Reactivating the water softener
41