Programming special functions
The following special functions can be
programmed where appropriate:
1. Extending the water inlet from 60
seconds to 120 seconds with level
control
If the inlet water pressure is lower than
35 psi (2.5 bar), too little water will flow
into the wash cabinet. By selecting the
extended water inlet, more water will
flow into the wash cabinet if there is low
water pressure (between 15 and 35 psi
[1.0 and 2.5 bar]).
Turn off the washer/disinfector.
Press DRAIN and DRYING touch
pads at the same time, while turning
on the machine. "P3" appears in the
display field.
Press DISINFECTION 93°C -10 MIN
touch pad.
”10” or ”11” will appear in the display.
10 = 60 seconds water intake
11 = 120 seconds water intake with
level control
Press DRYING touch pad to change
the display to ”11”.
Press START touch pad. ”SP” ap-
pears in the display field.
Press START again. The change is
now stored and the display field
clears.
2. Adding ML Neutralizer in the "In-
terim Rinse I" phase of the cycle
This special function should be se-
lected if signs of discoloration or corro-
sion on instruments occur, especially at
joints.
Turn off the washer/disinfector.
Press DRAIN and DRYING touch
pads at the same time, while turning
on the machine. "P3" appears in the
display field.
Press WASH 93°C touch pad.
”20” or ”21” will appear in the dis-
play field.
20 = without interim rinse I
21 = with neutralizing agent dis-
pensed in interim rinse I
Press DRYING touch pad to change
from 20 to 21 or vice versa.
Press START touch pad. "SP" ap-
pears in the display field.
Press START again. The change is
now stored and the display field
clears.
Programming special functions
38