Installation Verification
2.4 Essential
Performance
Essential Performance (EP) for surgical luminaires:
1.
Peak central illuminance (measured at 1m from the luminaire)
within the range from 40 klx to 160 klx.
2.
Peak central irradiance (measured at any distance along the
beam axis) less than 1000 W/m
2
.
3.
Provide white light to illuminate the operating field. (The apparent
whiteness of the light can be verified by visual observation.)
4.
Loss of illumination could result in delay of surgical procedure.
5.
In extreme cases of electromagnetic interference, the operator
may experience temporary loss of light function or camera
operation. This can be corrected by removing the source of
interference and power cycling the light if necessary.