ii
10043124
Operator Manual
Introduction
limiting the time of exposure at higher illumination levels. Extra care must
be taken when the light fields from multiple lightheads are overlapped on
the surgical site, since this condition creates a risk of too much heat.
An international standard for the safety of surgical lights established by the
International Electrotechnical Commission (IEC) sets minimum and
maximum levels of illumination and maximum levels of radiant heat that
can be emitted from a single surgical luminaire. The HarmonyAIR Surgical
Lighting System G Series has been designed to comply with this
international standard and to provide a wide range of illumination levels
while minimizing the potentially damaging infrared heat in the surgical
field.
The illumination level of surgical lights can be adjusted through several
intensity settings via conveniently located controls on either the wall-
mounted control center or the lighthead handle. The illumination level also
decreases as the pattern size increases. Maximum illuminance can reach
160 klx for the smallest pattern size of the lighthead and can be adjusted
by intensity control or pattern size control throughout the entire range
specified by the IEC.
Serious incidents that have occurred in relation to this medical device
should be reported to the manufacturer and competent authority in the
country where the incident occurred.
Manufactured by:
STERIS Corporation
2720 Gunter Park East
Montgomery, AL 36109 • USA
334-277-6660/800-444-9009
www.steris.com
Class 1 Equipment
Ordinary Equipment (enclosed equipment without protection from ingress
of water)
Equipment not suitable for use in the presence of a flammable anesthetic
mixture with air or oxygen or nitrous oxide.
Suitable for continuous operation.
The base language of this document is ENGLISH.
Any translations must be made from the base
language document.
STERIS Ireland Limited
IDA Business and
Technology Park
Tullamore
County Offaly
R35 X865
■
Ireland
Содержание HarmonyAIR G Series
Страница 1: ...OPERATOR MANUAL HarmonyAIR Surgical Lighting System G Series Rev G 10043124 ...
Страница 2: ......
Страница 8: ...vi 10043124 Operator Manual Table of Contents Healthcare Capital Equipment ...
Страница 10: ...ix Table of Contents Operator Manual 10043124 Healthcare Capital Equipment ...
Страница 24: ...2 6 10043124 Operator Manual Installation Verification Healthcare Capital Equipment ...
Страница 42: ...3 18 10043124 Operator Manual Operating Instructions Healthcare Capital Equipment ...
Страница 46: ...4 4 10043124 Operator Manual Cleaning The Equipment Healthcare Capital Equipment ...
Страница 54: ...7 2 10043124 Operator Manual Replacement Parts Healthcare Capital Equipment ...
Страница 56: ...8 2 10043124 Operator Manual Waste Disposal Guidelines Healthcare Capital Equipment ...