Operator Manual
Operating Instructions
•
A given lighthead’s surgeon's control buttons adjust the intensity
level for that lighthead.
Important: Avoid control faults. Do not continuously press and
hold wall control buttons for more than 40 seconds.
3.1.2 Wall Control Unit
Refer to
.
1.
Press the selected
+
button to activate the corresponding
lighthead.
•
Pressing the
+
button repeatedly incrementally increases the
intensity of light from the lighthead (until reaching maximum
intensity).
•
Pressing the
-
button repeatedly reduces light intensity until
reaching a minimum level. Continuing to press
-
button (for
2 seconds) at any intensity level causes the lighthead to shut
off. (Pressing that lightheads
+
button restores light to the
lighthead at previous level.) Refer to
for a
summary of wall control intensity functions.
2.
Once the lighthead is selected, press either the button to
increase light intensity level (
+
), or the button to decrease light
intensity level (
-
).
WARNING – EXPLOSION
HAZARD: Do not use this
lighting system in the
presence of flammable
anesthetics.
!
Table 3-1. HarmonyAIR Surgical Lighting System G Series Intensity Control
Module
State
Module
Intensity
Button
Pressed
Duration of
Switch Press
Operation
OFF
N/A
+
Any
Module turns ON at stored intensity setting
ON
Any except
highest
+
Any
Module increases to next higher intensity setting
ON
Highest
+
Any
Module remains at highest intensity setting
OFF
N/A
-
Any
Module remains OFF
ON
Lowest
-
Less than
2 seconds
Module remains at lowest intensity setting
ON
Lowest
-
2 seconds or longer
Module initially remains at lowest intensity
setting and after two seconds turns OFF
ON
Any except
lowest
-
Less than
2 seconds
Module decreases to next lower intensity setting
ON
Any except
lowest
-
2 seconds or longer
Module initially decreases to next lower intensity
setting and after two seconds turns OFF
Either
Any
Either
40 seconds or
longer
Switch Fault
: Depending upon the switch
pressed and the initial conditions of the G Series
module state and intensity, the control of the G
Series module follows one of the sequences
listed above. After 40 seconds, the switch is
disabled. All other switches remain operational.
Full control of G Series module intensity remains
available at the lighthead user interface.
Содержание HarmonyAIR G Series
Страница 1: ...OPERATOR MANUAL HarmonyAIR Surgical Lighting System G Series Rev G 10043124 ...
Страница 2: ......
Страница 8: ...vi 10043124 Operator Manual Table of Contents Healthcare Capital Equipment ...
Страница 10: ...ix Table of Contents Operator Manual 10043124 Healthcare Capital Equipment ...
Страница 24: ...2 6 10043124 Operator Manual Installation Verification Healthcare Capital Equipment ...
Страница 42: ...3 18 10043124 Operator Manual Operating Instructions Healthcare Capital Equipment ...
Страница 46: ...4 4 10043124 Operator Manual Cleaning The Equipment Healthcare Capital Equipment ...
Страница 54: ...7 2 10043124 Operator Manual Replacement Parts Healthcare Capital Equipment ...
Страница 56: ...8 2 10043124 Operator Manual Waste Disposal Guidelines Healthcare Capital Equipment ...