4-5
764333-766
2. Check table casters and floor locks for any
accumulated debris and clean as follows:
a.
Ensure floor locks are properly engaged
(approximately 60% of table weight is removed
from casters when equipped with self-
compensating floor lock system).
b.
Hold cleaner spray can 6-8" (152-203 mm) from
caster and spray cleaning fluid liberally on the
caster.
c.
Wipe caster with a cloth, dampened with water,
to remove cleaning fluid and debris. Ensure
items such as suture, oils and floor wax are
removed.
d.
Perform
Steps b and c
for remaining three
casters.
3. Articulate table through all movements and clean
all additional surfaces exposed during these
articulations as follows:
a.
Hold spray cleaner can 6-8" (152-203 mm) from
surface to be cleaned and spray cleaning fluid
liberally on surface.
b.
Wipe sprayed surfaces with a clean, lint-free
cloth, dampened with water to remove cleaning
fluid.
c.
Wipe cleaned surfaces again with a clean,
damp, lint-free cloth to remove any remaining
residue.
4. Turn hand control
OFF
when finished with cleaning
procedure.
5.
Perform
Steps 5 and 6
.
4.4 BATTERY CHARGING PROCEDURE
Important:
Batteries are recharged automatically as
long as the Main Power Switch is set to ON (lighted
green) and the ac power is connected to the STERIS
4085 General Surgical Table. The PLUG icon appears
on Power Panel LED Display (see Figure 4-1)
indicating ac power and diode bar indicates battery
condition.
Figure 4-1. INTELLIPOWER Panel
LED Display (Typical)
NOTE: Press
OFF
button on Primary Hand Control
when table is idle to ensure batteries are not depleted.
Significant power is used in Stand-By Mode.
STERIS 4085 General Surgical Table is equipped with
INTELLIPOWER
®
dual power system. The tables are
powered either by facility power or by internal
batteries. Motor/control batteries* require recharging
on a periodic basis depending on frequency of table
usage. If table is to be operated on battery power, note
the following:
Bar graph description for 120 V tables:
• When table is plugged into facility ac and Main
Power Switch is ON, diode bar shows battery charge
state. This battery charge state is also indicated by
icon on primary hand control.
• When table is on battery power and primary hand
control is active (on), diode bar shows battery charge
state. This battery charge state is also indicated by
icon on primary hand control.
Bar graph description for 220 V tables:
• When table is plugged into facility ac and Main
Power Switch is ON, diode bar shows battery charge
state. This battery charge state is also indicated by
icon on primary hand control.
• When table is on battery power and primary hand
control is active (on), diode bar is off. Battery charge
state is indicated by icon on primary hand control.
* All motor/control batteries are a sealed, lead-acid gel
electrolyte-type, with a nominal life of four years.
WARNING – PERSONAL INJURY AND/OR
EQUIPMENT DAMAGE HAZARD: Repairs
and adjustments to this equipment must be
made only by STERIS or STERIS-trained
service personnel. Maintenance performed
by unqualified personnel or installation of
unauthorized parts could cause personal
injury, result in improper equipment
performance, invalidate the warranty, or
result in costly damage. Contact STERIS
regarding service options.
WARNING – TRIPPING HAZARD: Route the
power cord to the receptacle to minimize
being tripped over by personnel in the area.
Label 2
ORANGE
GREEN
Label 1
CHARGE
BATTERY
Содержание 4085
Страница 1: ...MAINTENANCE MANUAL STERIS 4085 General Surgical Table 07 01 10 P764333 765 ...
Страница 3: ...OPERATOR MANUAL STERIS 4085 General Surgical Table 04 15 09 P150832 639 ...
Страница 61: ...MAINTENANCE MANUAL Maintenance Instructions STERIS 4085 General Surgical Table 07 01 10 P764333 766 ...
Страница 131: ......
Страница 161: ...32 764333 767 Figure 5 4085 Surgical Table Hydraulics 1 of 2 4 3 11 12 8 7 24 18 19 22 23 15 14 9 10 6 5 See 2 of 2 1 2 ...
Страница 172: ......
Страница 173: ......
Страница 174: ......
Страница 175: ......
Страница 176: ......
Страница 177: ......