3-4
150832-639
Operator Manual
Installation Instaructions
12.
Install leg section as follows:
NOTE: The STERIS 4085 General Surgical Table is equipped with Hi-
Lock™ locking mechanism enabling quick and easy removal/
installation of the leg section with a safety lock feature.
a.
Using hand control, position table top seat section to
horizontal by pressing
LEG UP
button (refer to
S
ECTION
4.2,
H
AND
C
ONTROL
)
.
b. Insert both leg section installation rods into seat frame until
fully engaged. Listen for "click" sound to indicate correct
installation (see Figure 3-8).
NOTE: The leg section should slide easily into the back frame
(seat section). Do not force.
c.
Pull leg section (both sides) to ensure correct installation.
This completes the STERIS 4085 General Surgical Table
installation. Refer to
for Operating Instructions
.
Figure 3-3. Connect Hand Control to STERIS 4085 General Surgical Table (Typical)
Control
Panel
Protective
Cover
Click
Содержание 4085