7-2
764333-766
Precautions must be taken when servicing the
hydraulic system. Working in the field is not the ideal
situation, but being aware of the need for maintaining
system cleanliness, using common sense and good
practices, and being dedicated to contamination
control should lead to success. However, if
contamination is suspected, purge system as indicated
in
7.2.2 Work Place
The Operating Room would be the ideal place to
perform hydraulic service, but a maintenance area or
some other “back room” is the more probable location.
Avoid, at all costs, working in or next to a room where
any woodworking is carried out. Almost as bad is an
area where metal work, especially grinding, is done.
The airborne particles generated by these or other
similar operations are devastating and uncontrollable.
If you have no choice, at least ensure no such activities
have been performed there for several hours and none
occur during servicing.
Try to work in a place with no nearby doors through
which people are continually passing. The less people-
traffic around you, the better. Stay away from heating
and cooling vents that also put particulates into the air.
7.2.3 Tools
Tools, gauges, hydraulic plugs, caps, etc., must be
extremely clean during servicing. Clean tools with a
lint-free cloth if possible. Hands and fingernails must
also be clean when handling hydraulics. Scrub
fingernails with a brush.
Always place tools and parts on a clean surface as work
is being done. And, of course, no smoking, eating or
drinking when working with hydraulics.
7.2.4 General Procedures
Wipe all components to be serviced, and areas adjacent
to them, with a lint-free cloth.
•
Breaking connections -
If connection is to remain
broken for any length of time, both ends of connection
must be plugged or capped with clean parts. If this is
impractical, at least cover open port with a lint-free
cloth. A rubber band can be used to hold cloth in place.
•
Removing components -
When a component is
removed, place it on, and cover it with, a lint-free
towel.
•
Handling parts -
Try to handle all hydraulic
components by touching only those surfaces which
do not interact with hydraulic system oil.
•
Assembling -
Examine all surfaces coming into
contact with hydraulic system oil (wetted surfaces).
Any visible particles (i.e., 0.002 inch or more in
diameter) must be removed. Also examine O-rings
and O-ring seats for scratches and/or tears, as well
as contamination.
Be careful any critical surfaces do not unnecessarily
touch anything during assembly.
Ensure all fittings, valves, etc., are torqued per
specifications.
•
Completing work -
Clean external surfaces of work
area as before. Any leaks are more easily detected
if all areas are clean before checking.
• Perform sensor calibration after servicing
articulating components.
When the work is complete, verify the table functions
normally ... not only the function that was repaired,
but all functions.
7.3 TABLE COLUMN COVER
ASSEMBLY
To gain access to most of the STERIS 4085 General
Surgical Table hydraulic and electrical components,
the Table Column Cover (or shroud) and/or Table
Base Cover must be removed. Remove the plastic or
stainless-steel Table Column shroud as follows:
1. Lower table for easier access.
2. Remove and save six top cover screws
3. Raise table to highest position to gain access to
lower cover screws.
4. Remove and save four lower cover screws and
washers (see Figure 7-2).
5. For stainless-steel version, pry open covers with flat
screwdriver. For plastic version, gently pull apart.
6. Remove and set safely aside in place to prevent
damage.
NOTE: To aid in loosening RTV between column covers
and plastic base cover for future procedures, spray Dry
Lubricant Spray NG 202 (P752870-091) along the
bottom edge before applying the RTV.
Figure 7-1. Table Column Top Cover Screws
w
Screw
Содержание 4085
Страница 1: ...MAINTENANCE MANUAL STERIS 4085 General Surgical Table 07 01 10 P764333 765 ...
Страница 3: ...OPERATOR MANUAL STERIS 4085 General Surgical Table 04 15 09 P150832 639 ...
Страница 61: ...MAINTENANCE MANUAL Maintenance Instructions STERIS 4085 General Surgical Table 07 01 10 P764333 766 ...
Страница 131: ......
Страница 161: ...32 764333 767 Figure 5 4085 Surgical Table Hydraulics 1 of 2 4 3 11 12 8 7 24 18 19 22 23 15 14 9 10 6 5 See 2 of 2 1 2 ...
Страница 172: ......
Страница 173: ......
Страница 174: ......
Страница 175: ......
Страница 176: ......
Страница 177: ......