2-4
150832-639
Operator Manual
Important User Information
2.3 Patient Safety
Straps
Adjust the patient safety straps such that they restrain the patient
according to currently accepted patient restraining practices, keeping
in mind the Trendelenburg/Reverse Trendelenburg, back, lateral tilt
and height adjustment.
2.4 General
Description
The STERIS 4085 General Surgical Table (see Figure 2-3):
•
is a mobile, electrohydraulically operated surgical table
specifically designed to provide the complete patient positioning
flexibility required for modern surgical care facilities.
•
features powered lateral tilt, Trendelenburg/Reverse
Trendelenburg and adjustable height functions.
•
is designed to safely function with patients not exceeding the
maximum patient weight limit (patient in normal orientation) of
1100 lb (499 kg) without patient posturing or tabletop slide
(except raise/lower); 1000 lb (454 kg) full patient posturing
(except tabletop slide); 600 lb (272 kg) full patient posturing
(including tabletop slide).
•
is constructed of aluminum alloy, stainless steel and other high
quality materials.
•
is equipped with a large sliding table top providing unrestricted
radiological access without patient reversing.
•
is powered by either internal battery or facility electric through use
of the patent-pending INTELLIPOWER
®
dual power system.
The table accepts positioning commands from three sources:
1.
A
hand control
(which includes “FLEX”, "REFLEX", "SLIDE HEAD",
"SLIDE FOOT", "STOP", "TREND”, "REVERSE TREND", “HEIGHT
UP”, "HEIGHT DN", “TILT LEFT”, "TILT RIGHT", "BACK DN", "BACK
UP", "LEG DN" , "LEG UP", "KIDNEY UP", "KIDNEY DN", "LEVEL",
"Normal and Reverse Orientation with LED" and "FLOOR LOCK".
Refer to Figure 4-3 for control feature locations.
2.
An optional physician-controlled
foot control
(which includes
Trendelenburg Tilts, Lateral Tilts and Height functions). Refer to
.
3.
An
Auxiliary Control System
permits table operation in the event
of primary control malfunction (refer to
).
NOTE: The Head Section is manually adjustable.
* SAE Measurements are approximate based on metric dimensions.
WARNING – PERSONAL
INJURY HAZARD: Failure to
keep the patient properly
secured with the patient
safety straps at all times could
result in death or serious
injury
.
CAUTION – POSSIBLE
EQUIPMENT DAMAGE:
Appropriate components of
this surgical table have been
tested and found in
compliance with IEC 60601-1-
2:2001, Medical Electrical
Equipment – Part 1; General
Requirements for Safety;
Electromagnetic Compatibility
(EMC). There is, however, a
potential for electromagnetic
or other interference between
this equipment and other
devices. Should user
experience interference,
relocate this device or
minimize the use of the
affected equipment while this
device is in operation.
Содержание 4085
Страница 1: ...MAINTENANCE MANUAL STERIS 4085 General Surgical Table 07 01 10 P764333 765 ...
Страница 3: ...OPERATOR MANUAL STERIS 4085 General Surgical Table 04 15 09 P150832 639 ...
Страница 61: ...MAINTENANCE MANUAL Maintenance Instructions STERIS 4085 General Surgical Table 07 01 10 P764333 766 ...
Страница 131: ......
Страница 161: ...32 764333 767 Figure 5 4085 Surgical Table Hydraulics 1 of 2 4 3 11 12 8 7 24 18 19 22 23 15 14 9 10 6 5 See 2 of 2 1 2 ...
Страница 172: ......
Страница 173: ......
Страница 174: ......
Страница 175: ......
Страница 176: ......
Страница 177: ......