High Voltage
High voltages in excess of 11.5kV are present within the amplifier. However, the equipment is
designed so that personnel cannot come into contact with high voltage circuits unless the covers are
removed.
Access inside the amplifier is restricted to suitably skilled engineers.
RF Radiation
Exposure of the human body to microwave radiation can constitute a hazard (ANSI/IEEE C95-1-
1992). Personnel must be protected from microwave energy produced by the TWT.
All RF connectors must be correctly fitted before operation, so that there is no leakage of RF energy.
The amplifier must not be operated unless the RF output connection is correctly terminated. It is
particularly hazardous to look into open waveguides, coaxial feeders, or transmitter antennae when
the amplifier is operating.
Input Drive
Damage to the amplifier may occur if the RF input drive level or the input frequency are outside the
limits given in the specification for the model (see section 1.1).
Environmental
•
Do not block, or poke objects through, the amplifier ventilation slots.
•
Do not spray aerosol cleaners directly on to the amplifier surfaces when cleaning.
•
or unsafe surface.
pped over or damaged by furniture or movable
equipment.
•
STA
provided if
necessary.
•
EEC Directive 89/336/EEC Electromagnetic compatibility.
•
EEC Directive 73/23/EEC Low voltage directive.
•
CETLus listed to ANSI/UL 60950-1-2002 & CAN/CSA22.2 No. 60950-1-03.
e2v technologies’ quality management system is certificated to the requirements of ISO9001:2000.
Do not operate the amplifier on an unstable
•
Ensure that all cables cannot be walked on, tri
•
Do not place objects inside the waveguide.
Ensure exhaust air is not re-circulated.
NDARDS COMPLIANCE
Stellar amplifiers comply with the following requirements. Further details may be
© e2v technologies (uk) limited 2010
DAS768892AA
Version
1,
page
3